The prevalence of pacemaker patients among the general population and of conducted energy devices (CED) for law enforcement and self-defence is increasing. Consequently, the question whether cardiac pacemaker (CPM) patients are on particular risk becomes increasingly important. The risk of Taser X26 electric interference with implanted CPM has been investigated by numerical simulation at MRI-based anatomical models of CPM patients with devices implanted at conventional sites (left pectoral, right pectoral and abdominal) and with the monopolar CPM electrode placed at the ventricular apex. In spite of 10fold higher peak voltages the different coupling conditions make Taser-induced CPM interference voltages lower than those caused by external cardiac defibrillators. It is shown that electric interference considerably depends on ECD electrode orientation. The most unfavourable conditions are encountered with ECD electrodes aligned with the line from the CPM electrode tip to CPM can (EPC line). It could be shown that worst case interference voltages of monopolar pacemakers of any kind of implantation remain below the pulse immunity level as defined in safety standards of implantable cardiac pacemakers and of cardioverter defibrillators. However, interference voltages exceed CPM sensing thresholds. Therefore, capturing should be expected at Taser X26 contact mode application at any position at the upper part of the body including the abdomen, both at frontal and dorsal positions.
机译:在一般人群中,用于执法和自卫的起搏器患者和传导能量设备(CED)的患病率正在上升。因此,心脏起搏器(CPM)患者是否有特殊风险的问题变得越来越重要。 Taser X26电干扰植入CPM的风险已通过在基于MRI的CPM解剖模型中进行数值模拟研究,该模型在常规部位(左胸,右胸和腹部)植入了器械并将单极CPM电极置于心室的CPM患者顶尖。尽管峰值电压高出10倍,但不同的耦合条件使Taser引起的CPM干扰电压低于由外部心脏除颤器引起的干扰电压。结果表明,电干扰在很大程度上取决于ECD电极的方向。与从CPM电极头到CPM罐的线(EPC线)对齐的ECD电极遇到最不利的条件。可以证明,任何植入类型的单极起搏器在最坏情况下的干扰电压都保持在可植入心脏起搏器和心脏复律除颤器安全标准所定义的脉冲抗扰度以下。但是,干扰电压超过CPM感应阈值。因此,在Taser X26接触模式应用中,应在包括腹部在内的身体上部任何位置的正面和背面位置进行捕捉。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号