Influence of time on primary stability of dental implants placed with osteotomes due to the elastic properties of peri-implant bone

机译:由于植入物周围骨的弹性,时间对带有截骨的牙科植入物的初步稳定性的影响

获取原文
           

摘要

Background/purposeThe use of osteotomes as a technique for densification and expansion of the residual ridge is one of the most widely used procedures to achieve adequate peri-implant bone quantity and density. The aim of this study was to evaluate the influence of time and the elastic deformation of peri-implant bone in the primary stability of implants placed using osteotomes.Materials and methodsIn each of 10 fresh fragments of cow rib, two implants were placed using osteotomes. The insertion torque and initial implant stability quotient (ISQ) values were measured. In the control implants, the immediate removal torque was measured, while in the test implants, after 15?min of placement, ISQ values were measured again, and the removal torque was measured.ResultsThere were significant differences between the ISQ values and between the insertion torque and removal torque at 15?min. The ISQ values (perpendicular/parallel) increased between the initial moment (64.4?±?9/70.3?±?5.9) and 15?min (66/71.4?±?6.4). The removal torque at 15?min (12.4?±?5.8) was lower than the insertion torque (15.9?±?5.9). Compression of the trabeculae in contact with the implant placed using osteotomes was observed, as well as a greater number of trabecular fractures in the implants placed using conventional drilling.ConclusionThere is an increase in ISQ values of dental implants placed using osteotomes after 15?min of placement.
机译:背景/目的使用截骨术作为增厚和扩大残余and的技术是获得适当的种植体周围骨量和密度的最广泛使用的方法之一。这项研究的目的是评估时间和种植体周围骨的弹性变形对使用截骨术放置的植入物的初步稳定性的影响。材料和方法在10个新鲜的牛肋骨碎片中,分别使用截骨术放置两个植入物。测量插入扭矩和初始植入物稳定性商(ISQ)值。在对照种植体中,测量立即去除扭矩,而在测试种植体中,放置15分钟后,再次测量ISQ值,并测量去除扭矩。结果ISQ值与插入之间存在显着差异。扭矩和拆卸扭矩为15?min。 ISQ值(垂直/平行)在初始时刻(64.4?±?9 / 70.3?±?5.9)和15?min(66 / 71.4?±?6.4)之间增加。 15μmin时的去除扭矩(12.4≤±5.8)低于插入扭矩(15.9≤±5.9)。观察到与使用截骨刀放置的植入物接触的小梁受压,以及使用常规钻孔放置的植入物中的小梁骨折数量较多。结论在15分钟后,使用截骨术放置的牙科植入物的ISQ值增加。放置。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号