System Accuracy of Blood Glucose Monitoring Devices According to the Current and Proposed ISO 15197 Standards

机译:根据当前和拟议的ISO 15197标准的血糖监测设备的系统精度

获取原文
           

摘要

D aily diabetes management requires blood glucose monitoring systems (BGMS), which provide accurate readings, such as OneTouch ? Verio? meter devices and test strips (LifeScan, Inc., Milpitas, CA). Two recent studies 1,2 that evaluated several BGMSs reported that these systems did not, in all cases, meet the current and the draft revision International Organization for Standardization (ISO) 15197 standards. 3,4 This contrasts with previous LifeScan-sponsored studies evaluating the accuracy of our three BGMSs (A, B, and C), which reported that these systems met the current ISO 15197 criteria, 5 and were particularly accurate at low blood glucose levels. 6 A reanalysis of the same data, 5,6 presented here, further shows that the aforementioned three BGMSs meet the current and proposed ISO 15197 criteria for system accuracy (refer to Table 1).
机译:日常糖尿病管理需要血糖监测系统(BGMS),该系统可提供准确的读数,例如OneTouch?维里奥?仪表设备和测试条(加利福尼亚州米尔皮塔斯市的LifeScan,Inc.)。最近的两项研究1,2评估了多个BGMS,报告称这些系统在所有情况下均未达到国际标准化组织(ISO)15197的当前和修订草案。 3,4与之前由LifeScan赞助的评估我们三个BGMS(A,B和C)准确性的研究形成对比,后者报告这些系统符合当前的ISO 15197标准5,并且在低血糖水平下特别准确。 6对此处相同数据(5,6)的重新分析进一步表明,上述三个BGMS符合当前和建议的ISO 15197系统精度标准(请参阅表1)。