首页> 外文期刊>Journal of Diabetes Science and Technology >Accuracy and Reliability of the Nova StatStrip? Glucose Meter for Real-Time Blood Glucose Determinations during Glucose Clamp Studies
【24h】

Accuracy and Reliability of the Nova StatStrip? Glucose Meter for Real-Time Blood Glucose Determinations during Glucose Clamp Studies

机译:Nova StatStrip的准确性和可靠性?血糖仪,用于在血糖钳位研究中实时测定血糖

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Aims/Hypothesis: The Andres clamp technique, which requires accurate and timely determination of glucose, utilizes the Beckman or Yellow Springs Instruments (YSI) glucose analyzers. Both instruments require maintenance, a dedicated operator, preparation of a plasma sample, and a duplicate measurement that takes ≥2 minutes. The Nova StatStrip glucose meter was evaluated for accuracy, reliability, and near-real-time availability of glucose. Methods: Blood samples from 24 patients who underwent 6-hour clamp studies and 12 patients who had a standardized meal tolerance test (SMT) were measured. Specimens were analyzed simultaneously and immediately upon collection by Beckman, YSI, and Nova. Results: Of 1004 data pairs for the Nova device versus Beckman, the Nova data points ranged from 32 to 444, while Beckman ranged from 42 to 412. The coefficient for the slope of Beckman versus Nova was 1.009 ( r = 0.978). Using error grid analysis, the number and percentage of values for Nova were 976 (97.2%) in the A zone and 28 (2.8%) in the B zone. Of 399 data pairs for the Nova device versus YSI, the Nova data points ranged from 46 to 255, whereas YSI ranged from 47 to 231. The coefficient for the slope of YSI versus Nova was 1.023 ( r = 0.989). All Nova readings fell in the A zone. Time required for final reading, in duplicate, was 15 seconds for Nova and 120–180 seconds for Beckman and YSI. Conclusions: The simplicity of Nova and its reliability, accuracy, and speed make it an acceptable replacement device for Beckman and YSI in the conduct of clamps, especially when perturbations require rapid glucose determination.
机译:目的/假设:安德列斯钳制技术要求准确,及时地确定葡萄糖,它利用贝克曼或黄泉仪器(YSI)的葡萄糖分析仪。两种仪器都需要维护,专职操作人员,血浆样品的制备以及重复测量的时间≥2分钟。对Nova StatStrip血糖仪进行了准确性,可靠性和近实时葡萄糖可用性评估。方法:对24名进行了6小时钳制研究的患者和12名进行了标准膳食耐受性测试(SMT)的患者的血样进行了测量。 Beckman,YSI和Nova会在收集样本后立即对其进行同时分析。结果:在Nova设备与Beckman的1004个数据对中,Nova数据点的范围为32至444,而Beckman从42到412。Beckman与Nova的斜率系数为1.009(r = 0.978)。使用误差网格分析,A区的Nova值的数量和百分比为976(97.2%),B区为28(2.8%)。 Nova设备与YSI的399个数据对中,Nova数据点的范围为46至255,而YSI的范围为47至231。YSI与Nova的斜率系数为1.023(r = 0.989)。新星的所有读数都落在A区。最终阅读所需的时间(一式两份)对于Nova来说是15秒,对于Beckman和YSI是120-180秒。结论:Nova的简单性,可靠性,准确性和速度使其成为贝克曼(Beckman)和YSI的钳夹操作可接受的替代设备,尤其是在扰动需要快速测定葡萄糖的情况下。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号