PurposeThe Nova StatStrip (R) Glucose Hospital Meter System (Nova Biomedical, Waltham, MA, USA) is United States Food and Drug Administration approved for point-of-care use in critically ill patients, but its use during cardiac surgery has not been evaluated. In this study, we compare glucose values obtained during cardiac surgery by StatStrip (R) with values obtained by a blood gas analyzer.MethodsBlood glucose concentrations were analyzed in 121 patients by the StatStrip point- of-care test (POCT) glucose monitor and the GEM (R) Premier (TM) 3000 blood gas analyzer (Instrumentation Laboratory Company, Bedford MA, USA). Arterial blood samples were taken at baseline (before surgery), before cardiopulmonary bypass (CPB), during early and late CPB, and 30 min after CPB. Accuracy of the StatStrip glucometer was analyzed using the Clinical and Laboratory Standards Institute (CLSI) POCT12-A3 criteria (criterion 1; 95% of samples should be 0.66 mMolL(-1) of reference glucose values 5.5 mMolL(-1), criterion 2; 98% of samples should be 0.83 mMolL(-1) of reference glucose values 4.1 mMolL(-1)).ResultsThe accuracy of StatStrip glucose measurements at baseline (99%, 100%) and before CPB (95%, 98%), but not during (early: 84%, 97%; late: 83%, 96%) and after (92%, 100%) CPB, satisfied the CLSI POCT12-A3 criteria.ConclusionArterial blood glucose measurement by StatStrip was accurate before CPB, but lacked accuracy during and after CPB. Glucose values should be interpreted with caution when intensive glucose control protocols are being used during cardiac surgery.Trial registrationClinicalTrials.gov (NCT02729064); registered 5 April, 2016.
机译:Purposethe Nova Statstrip(R)葡萄糖医院仪表系统(Nova Biomedical,沃尔瑟姆,MA,美国)是美国食品和药物管理局批准用于危重病患者的护理点,但它在心脏手术期间的使用尚未评估。在这项研究中,我们通过Statstrip(R)在心脏手术期间获得的葡萄糖值通过血气分析仪获得的值。通过StatsTip的治疗性测试(POCT)葡萄糖监测器和葡萄糖监测器中的121名患者分析了葡萄糖浓度的葡萄糖浓度。 GEM(R)总理(TM)3000血气分析仪(仪器实验室公司,Bedford MA,USA)。在基线(手术前),在心肺旁路(CPB),在早期和晚期CPB期间,CPB后30分钟拍摄动脉血样。使用临床和实验室标准研究所(CLSI)POCT12-A3标准进行分析Statstrip Glucometer的准确性(标准1; 95%的样品应为0.66 mmoll(-1)参考葡萄糖值5.5 mmoll(-1),标准2 ; 98%的样品应为0.83 mmoll(-1)参考葡萄糖值4.1 mmoll(-1))。基线(99%,100%)和CPB之前的STATSTP葡萄糖测量的准确度(95%,98%)但是,但在(早期:84%,97%;晚期:83%,96%)和之后(92%,100%)CPB,满足CLSI POCT12-A3标准。CPB之前,CLSIIP血糖测量是准确的,但在CPB期间和之后缺乏准确性。在心脏手术期间使用密集的葡萄糖控制方案时,应谨慎地解释葡萄糖值.TirdritedClinicalTrials.gov(NCT02729064);注册2016年4月5日。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号