Differences in erythrocyte sedimentation rates using a modified Westergren method and an alternate method

机译:使用改良的Westergren方法和替代方法的红细胞沉降率差异

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Introduction Worldwide laboratories have adopted the use of modified or alternate methods for measurement of the erythrocyte sedimentation rate (ESR). The iSED from Alcor Scientific is a novel, alternate ESR method based on photometric aggregometry which offers improved operator safety and reduced analysis time. This study evaluated the diagnostic utility of the iSED in a South African patient population with a range of inflammatory disorders. Methods We compared the iSED with the predicate modified Westergren method (StaRRsed, Mechatronics, Zwaag, the Netherlands) measured at 60?minutes. Analysis was performed on Ksub2/subEDTA samples at three ESR measurement ranges (80?mm/h) in 120 pediatric and adult inpatients and outpatients over a 2‐week period. Precision, stability, and carryover were performed in accordance with the revised International Council for Standardisation in Haematology guidelines. Results The iSED demonstrated acceptable imprecision with minimal carryover (2.86%). The correlation coefficients at the 3 ESR measurement ranges were r =?0.58, r =?0.71, and r =?0.56, respectively. The y‐intercepts were ?10.74 (CI ?29.17 to 7.69), ?5.95 (CI ?18.60 to 6.69) and 246.05 (CI 591.42‐99.31). This indicated a difference of a constant nature with an overall mean difference of 7.99?mm/h (CI 5.87‐10.13) ( P Conclusion This study demonstrated differences in ESR results, predominantly at extremes of the analytical range, using an alternate method. Careful consideration and performance monitoring of these novel methods are advised.
机译:简介世界各地的实验室已采用改进或替代方法来测量红细胞沉降速率(ESR)。 Alcor Scientific的iSED是一种新颖的,基于光度聚集法的替代ESR方法,可提高操作员的安全性并减少分析时间。这项研究评估了iSED在具有多种炎症性疾病的南非患者人群中的诊断效用。方法我们将iSED与谓词改进的Westergren方法(StaRRsed,Mechatronics,Zwaag,荷兰)进行了比较,在60分钟时进行了测量。在2个星期的时间内,对120个儿科和成人住院患者和门诊患者的三个ESR测量范围(80?mm / h)的K 2 EDTA样品进行了分析。根据修订的国际血液学标准化理事会的指南进行精密度,稳定性和残留分析。结果iSED表现出可接受的不精确度,残留率极低(2.86%)。在3个ESR测量范围内的相关系数分别为r =?0.58,r =?0.71和r =?0.56。 y截距为10.74(CI为29.17至7.69),5.95(CI为18.60至6.69)和246.05(CI 591.42-99.31)。这表明存在恒定性质的差异,总体平均差异为7.99?mm / h(CI 5.87-10.13)(P结论本研究表明,ESR结果存在差异,主要是在分析范围的极限处,使用了另一种方法。建议对这些新颖方法进行考虑和性能监控。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号