Background Blood gas analysis and blood lactate measurement have important roles in patient management. Point‐of‐care (POC) testing simplifies and provides rapid blood gas and lactate measurements. This study aimed to compare pH, pCOsub2/sub, pOsub2/sub, and lactate measurements between a POC device and a benchtop blood gas analyzer typically used in a hospital central laboratory, and to evaluate the inter‐device variability of the POC device. Methods A cross‐sectional study was conducted with a sample size of 100. Each sample was measured for pH, pCOsub2/sub, pOsub2/sub, and lactate using a Nova pHOx plus Lsup?/sup benchtop blood gas analyzer in the central laboratory and an i‐STATsup?/sup handheld POC device. The results of both devices were compared using Pearson or Spearman correlation coefficients and Bland‐Altman tests. Testing of the inter‐device variability was done by using three different i‐STATsup?/sup devices, and the results were compared statistically. Results Strong correlations were observed for all test results. In Bland‐Altman analysis, ≥95% of the results were within the limits of agreement, with the exception of lactate, which had only 93%. The results that were beyond the limits were primarily lactate levels 8?mmol/L. Biases between the benchtop analyzer and the i‐STATsup?/sup were not clinically significant, except pH. No significant inter‐device variability was observed between the i‐STATsup?/sup analyzers. Conclusion This comparison study of pH, pCOsub2/sub, pOsub2/sub, and lactate measurements between Nova pHOx plus Lsup?/sup and i‐STATsup?/sup analyzers showed good agreement. However, lactate measurement results 8?mmol/L on the i‐STATsup?/sup analyzer should be interpreted with caution.
机译:背景技术血气分析和血乳酸测量在患者管理中具有重要作用。即时护理(POC)测试可简化并提供快速的血气和乳酸测量。这项研究旨在比较POC设备和通常在医院中心实验室中使用的台式血气分析仪之间的pH,pCO 2 ,pO 2 和乳酸测量值,并评估POC设备的设备间差异。方法采用100份样本进行横断面研究。使用Nova pHOx加L法测量每个样本的pH,pCO 2 ,pO 2 和乳酸中央实验室的?台式血气分析仪和i‐STAT ?手持式POC设备。使用Pearson或Spearman相关系数和Bland-Altman检验比较了两种设备的结果。通过使用三种不同的i-STAT ?设备对设备间的可变性进行测试,并对结果进行统计比较。结果对于所有测试结果均观察到强相关性。在Bland-Altman分析中,≥95%的结果在一致的范围内,但乳酸盐除外,只有93%。超出限值的结果主要是乳酸水平> 8?mmol / L。台式分析仪和i-STAT ?之间的偏差在临床上无显着意义,但pH值除外。在i-STAT ?分析仪之间未观察到明显的设备间差异。结论Nova pHOx加L ?和i-STAT 之间的pH,pCO 2 ,pO 2 和乳酸测量的比较研究?分析仪显示出良好的一致性。但是,应谨慎解读i‐STAT ?分析仪上的乳酸测量结果> 8?mmol / L。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号