摘要

We developed a two-dimensional plot for viewing trueness that takes into account potential shift and variable quality requirements to verify trueness using certified reference material (CRM).MethodsGlucose, total cholesterol (TC), and creatinine levels were determined by two kinds of assay in two levels of a CRM. Available quality requirements were collected, codified, and sorted in an ascending order in the plot's header row. Centering on the mean of measured values from CRM, the “mean ± US CLIA '88 allowable total error” was located in the header of the leftmost and rightmost columns. Twenty points were created in intervening columns as potential shifts. Uncertainties were calculated according to regression between certified values and uncertainties of CRM, and positioned in the corresponding columns. Cells were assigned different colors where column and row intersected based on comparison of the 95% confidence interval of the percentage bias with each quality requirement.ResultsA glucose assay failed to meet the highest quality criteria, for which shift of +0.13–0.14 mmol/l was required. A TC assay met the quality requirement and a shift of ±0.03 mmol/l was tolerable. A creatinine assay also met the quality requirement but any shift was not tolerable.ConclusionThe plot provides a systematic view of the trueness of quantitative laboratory tests.
机译:我们开发了一个用于查看真实性的二维图,其中考虑了潜在的偏移和可变质量要求,以使用认证参考材料(CRM)验证真实性。方法葡萄糖,总胆固醇(TC)和肌酐水平通过两种检测方法确定。 CRM的两个级别。在地块的标题行中,收集可用的质量要求,将其编码并按升序排序。以CRM的测量值的平均值为中心,“平均值±US CLIA '88允许的总误差”位于最左边和最右边列的标题中。由于潜在的移动,在中间的列中创建了20个点。根据认证值和CRM的不确定性之间的回归来计算不确定性,并将其定位在相应的列中。根据对百分比偏差的95%置信区间与每种质量要求的比较,为细胞分配了不同的颜色,其中行和行相交。结果葡萄糖测定未能满足最高质量标准,其位移为+ 0.13-0.14 mmol / l是必需的。 TC分析符合质量要求,±0.03 mmol / l的变化是可以容忍的。肌酐测定法也符合质量要求,但任何变化都是不能容忍的。结论该图提供了定量实验室测试真实性的系统视图。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号