Evaluation of Different Disinfectants on the Performance of an On-Meter Dosed Amperometric Glucose-Oxidase-Based Glucose Meter

机译:不同消毒剂对基于流量计的安培葡萄糖氧化酶基葡萄糖测定仪性能的评估

获取原文
           

摘要

Background: Off-meter dosed photometric glucose-oxidase-based glucose meters have been reported to be susceptible to interference by hydrogen-peroxide-based disinfecting agents. The objective of this study was to determine if a single application of hydrogen-peroxide-containing Accel? wipe to disinfect an on-meter dosed amperometric glucose-oxidase-based glucose meter will influence its performance. Method: The performance of five on-meter dosed amperometric glucose-oxidase-based glucose meters was determined before and after disinfecting the devices with a single application of either CaviWipes? (14.3% isopropanol and 0.23% diisobutyl-phenoxy-ethoxyethyl dimethyl benzyl ammonium chloride) or Accel (0.5% hydrogen peroxide) wipes. Replicate glucose measurements were conducted before disinfecting the devices, immediately after disinfecting, and then 1 and 2 min postdisinfecting, with measurements in triplicate. Analysis was sequentially completed for five different meters. Results were analyzed by a two-way analysis of variance (Analyze-it software). Results: No clinical (<0.3 mmol/liter) or statistical differences ( p > .05) in glucose concentration were detected when the on-meter dosed amperometric glucose-oxidase-based glucose meters were disinfected with either CaviWipes or Accel wipes and measured immediately or 1 or 2 min postdisinfecting. No clinically significant difference in glucose concentration was detected between meters (<0.3 mmol/liter). Conclusion: The on-meter dosed glucose oxidase amperometric-based glucose meters are not analytically susceptible to interference by a single application of hydrogen-peroxide-containing Accel disinfectant wipes.
机译:背景:据报导,基于仪表外剂量的基于光度法的基于葡萄糖氧化酶的血糖仪容易受到基于过氧化氢的消毒剂的干扰。这项研究的目的是确定是否一次使用过氧化氢的Accel?擦拭以消毒基于流量计的定量安培型基于葡萄糖氧化酶的葡萄糖计会影响其性能。方法:在一次使用CaviWipes?消毒设备之前和之后,确定了五台基于流量计的剂量式安培型基于葡萄糖氧化酶的血糖仪的性能。 (14.3%异丙醇和0.23%二异丁基-苯氧基-乙氧基乙基二甲基苄基氯化铵)或Accel(0.5%过氧化氢)擦布。在设备消毒之前,消毒之后,消毒后1分钟和2分钟进行一式三份的重复葡萄糖测量。依次完成了对五个不同仪表的分析。通过双向方差分析(Analyze-it软件)分析结果。结果:当使用CaviWipes或Accel擦拭巾对按表剂量的安培型基于葡萄糖氧化酶的血糖仪进行消毒并立即进行测量时,未发现葡萄糖浓度的临床(<0.3 mmol / L)或统计学差异(p> .05)。或消毒后1或2分钟。在米之间(<0.3 mmol / L)未检测到临床上葡萄糖浓度的显着差异。结论:基于计量的基于葡萄糖氧化酶安培的血糖仪在分析上不易受到单次使用含过氧化氢的Accel消毒湿巾的干扰的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号