首页> 外文期刊>Journal of applied clinical medical physics / >Quantitative variations in texture analysis features dependent on MRI scanning parameters: A phantom model
【24h】

Quantitative variations in texture analysis features dependent on MRI scanning parameters: A phantom model

机译:纹理分析特征的定量变化取决于MRI扫描参数:幻影模型

获取原文
           

摘要

Objectives To evaluate the influence of MRI scanning parameters on texture analysis features. Methods Publicly available data from the Reference Image Database to Evaluate Therapy Response (RIDER) project sponsored by The Cancer Imaging Archive included MRIs on a phantom comprised of 18 25‐mm doped, gel‐filled tubes, and 1 20‐mm tube containing 0.25?mM Gd‐DTPA (EuroSpinII Test Object5, Diagnostic Sonar, Ltd, West Lothian, Scotland). MRIs performed on a 1.5?T GE HD, 1.5?T Siemens Espree (VB13), or 3.0?T GE HD with TwinSpeed gradients with an eight‐channel head coil included T1WIs with multiple flip angles (flip‐angle?=?2,5,10,15,20,25,30), TR/TE?=?4.09–5.47/0.90–1.35?ms, NEX?=?1 and DCE with 30° flip‐angle, TR/TE=4.09–5.47/0.90–1.35, and NEX?=?1,4. DICOM data were imported into an in‐house developed texture analysis program which extracted 41‐texture features including histogram, gray‐level co‐occurrence matrix (GLCM), and gray‐level run‐length (GLRL). Two‐tailed t tests, corrected for multiple comparisons ( Q values) were calculated to compare changes in texture features with variations in MRI scanning parameters (magnet strength, flip‐angle, number of excitations (NEX), scanner platform). Results Significant differences were seen in histogram features (mean, median, standard deviation, range) with variations in NEX ( Q =?0.003–0.045) and scanner platform ( Q Conclusion Significant differences were seen in many texture features with variations in MRI acquisition emphasizing the need for standardized MRI technique.
机译:目的评估MRI扫描参数对纹理分析特征的影响。方法由癌症影像档案馆赞助的可从参考图像数据库评估治疗反应(RIDER)项目获得的公开数据包括幻像上的MRI,该幻像由18个25mm掺杂凝胶填充试管和1个20mm试管(含0.25?)组成。 mM Gd-DTPA(EuroSpinII测试对象5,诊断声纳有限公司,西洛锡安,苏格兰)。在具有TwinSpeed梯度的1.5?T GE HD,1.5?T Siemens Espree(VB13)或3.0?T GE HD上进行的MRI,带有八通道头部线圈,包括具有多个翻转角度的T1WI(翻转角度== 2, 5,10,15,20,25,30),TR / TE?=?4.09–5.47 / 0.90–1.35?ms,NEX?=?1且DCE具有30°翻转角,TR / TE = 4.09–5.47 /0.90-1.35,并且NEX?=?1,4。将DICOM数据导入内部开发的纹理分析程序中,该程序提取了41种纹理特征,包括直方图,灰度共现矩阵(GLCM)和灰度游程(GLRL)。计算出经过两次校正的两尾t检验(Q值),以比较质地特征的变化与MRI扫描参数(磁体强度,翻转角,激发数(NEX),扫描仪平台)的变化。结果直方图特征(均值,中位数,标准差,范围)存在显着差异,NEX(Q =?0.003-0.045)和扫描仪平台存在差异(Q结论在许多纹理特征中均存在显着差异,MRI采集强调需要标准化的MRI技术。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号