首页> 外文期刊>Journal of applied clinical medical physics / >Dosimetric study of GZP6 60Co high dose rate brachytherapy source
【24h】

Dosimetric study of GZP6 60Co high dose rate brachytherapy source

机译:GZP6 60Co高剂量率近距离放射治疗源的剂量学研究

获取原文
获取外文期刊封面目录资料

摘要

The purpose of this study was to obtain dosimetric parameters of GZP6 60Co brachytherapy source number 3. The Geant4 MC code has been used to obtain the dose rate distribution following the American Association of Physicists in Medicine (AAPM) TG‐43U1 dosimetric formalism. In the simulation, the source was centered in a 50?cm radius water phantom. The cylindrical ring voxels were 0.1?mm thick for r ≤?1?cm, 0.5?mm for 1?cm??5?cm. The kerma‐dose approximation was performed for r >?0.75?cm to increase the simulation efficiency. Based on the numerical results, the dosimetric datasets were obtained. These results were compared with the available data of the similar 60Co high dose rate sources and the detailed dosimetric characterization was discussed.
机译:这项研究的目的是获得GZP6 60 Co近距离放射源3的剂量学参数。根据美国医学物理学会(AAPM),使用Geant4 MC代码获得剂量率分布TG‐43U1剂量形式学。在模拟中,源位于半径为50?cm的水体模型中。圆柱形环状体素的厚度为r≤≤1μΩcm为0.1μmm,1μcm≤5μcm为0.5μmm。 r>?0.75?cm时进行了比释比的近似计算,以提高仿真效率。基于数值结果,获得了剂量数据集。将这些结果与类似的 60 Co高剂量率来源的可用数据进行了比较,并讨论了详细的剂量学表征。