Abstract ObjectiveAnalytical dose calculation algorithms for Eclipse and Raystation treatment planning systems (TPS), as well as a Raystation Monte Carlo model are compared to corresponding measured point doses. MethodThe TPS were modeled with the same beam data acquired during commissioning. Thirty-five typical plans were made with each planning system, 31 without range shifter and four with a 5 cm range shifter. Point doses in these planes were compared to measured doses. ResultsThe mean percentage difference for all plans between Raystation and Eclipse were 1.51 ± 1.99%. The mean percentage difference for all plans between TPS models and measured values are ?2.06 ± 1.48% for Raystation pencil beam (PB), ?0.59 ± 1.71% for Eclipse and ?1.69 ± 1.11% for Raystation monte carlo (MC). The distribution for the patient plans were similar for Eclipse and Raystation MC with a P -value of 0.59 for a two tailed unpaired t -test and significantly different from the Raystation PB model with P = 0.0013 between Raystation MC and PB. All three models faired markedly better if plans with a 5 cm range shifter were ignored. Plan comparisons with a 5 cm range shifter give differences between Raystation and Eclipse of 3.77 ± 1.82%. The mean percentage difference for 5 cm range shifter plans between TPS models and measured values are ?3.89 ± 2.79% for Raystation PB, ?0.25 ± 3.85% for Eclipse and 1.55 ± 1.95% for Raystation MC. ConclusionBoth Eclipse and Raystation PB TPS are not always accurate within ±3% for a 5 cm range shifters or for small targets. This was improved with the Raystation MC model. The point dose calculations of Eclipse, Raystation PB, and Raystation MC compare within ±3% to measured doses for the other scenarios tested.
机译:摘要目的将Eclipse和Raystation治疗计划系统(TPS)的分析剂量计算算法以及Raystation Monte Carlo模型与相应的测量点剂量进行比较。方法使用调试期间获得的相同光束数据对TPS进行建模。每个计划系统制定了35个典型计划,其中31个不带换档器,四个带5厘米的换档器。将这些平面中的点剂量与测量的剂量进行比较。结果Raystation和Eclipse之间所有计划的平均百分比差异为1.51±1.99%。 TPS模型与测量值之间所有计划的平均百分比差异为Raystation笔形束(PB)为2.06±1.48%,Eclipse为0.59±1.71%,Raystation monte carlo(MC)为1.69±1.11%。 Eclipse和Raystation MC的患者计划分布相似,两个未配对t尾检验的P值为0.59,与Raystation MC和PB之间的P = 0.0013的Raystation PB模型显着不同。如果忽略了带5厘米范围移位器的计划,则这三个模型的表现都明显更好。使用5厘米范围的移位器进行的计划比较使Raystation和Eclipse之间的差异为3.77±1.82%。 TPS模型与测量值之间的5厘米范围移位器计划的平均百分比差异为:Raystation PB为±3.89±2.79%,Eclipse为±0.25±3.85%,而Raystation MC为1.55±1.95%。结论对于5厘米范围的移位器或小型目标,Eclipse和Raystation PB TPS都不能始终在±3%的范围内准确。 Raystation MC模型对此进行了改进。 Eclipse,Raystation PB和Raystation MC的点剂量计算与其他测试情况下的测量剂量相比在±3%之内。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号