VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests

机译:VMAT线性加速器的调试和质量保证:剂量控制和龙门速度测试

获取原文
获取外文期刊封面目录资料

摘要

In VMAT treatment delivery the ability of the linear accelerator (linac) to accurately control dose versus gantry angle is critical to delivering the plan correctly. A new VMAT test delivery was developed to specifically test the dose versus gantry angle with the full range of allowed gantry speeds and dose rates. The gantry-mounted IBA MatriXX with attached inclinometer was used in movie mode to measure the instantaneous relative dose versus gantry angle during the plan every 0.54 s. The results were compared to the expected relative dose at each gantry angle calculated from the plan. The same dataset was also used to compare the instantaneous gantry speeds throughout the delivery compared to the expected gantry speeds from the plan. Measurements performed across four linacs generally show agreement between measurement and plan to within 1.5% in the constant dose rate regions and dose rate modulation within 0.1 s of the plan. Instantaneous gantry speed was measured to be within 0.11 ° / s of the plan (1 SD). An error in one linac was detected in that the nominal gantry speed was incorrectly calibrated. This test provides a practical method to quality-assure critical aspects of VMAT delivery including dose versus gantry angle and gantry speed control. The method can be performed with any detector that can acquire time-resolved dosimetric information that can be synchronized with a measurement of gantry angle. The test fulfils several of the aims of the recent Netherlands Commission on Radiation Dosimetry (NCS) Report 24, which provides recommendations for comprehensive VMAT quality assurance.PACS number(s): 87.55.Qr
机译:在VMAT治疗中,直线加速器(直线加速器)准确控制剂量与机架角度的能力对于正确交付计划至关重要。开发了一种新的VMAT测试装置,以在允许的龙门速度和剂量率的整个范围内专门测试剂量与龙门角度的关系。在电影模式下,将带有倾斜仪的安装在机架上的IBA MatriXX安装在机架上,每0.54 s测量一次相对剂量与机架角度的关系。将结果与根据计划计算的每个机架角度的预期相对剂量进行比较。同一数据集还用于比较整个交付过程中的瞬时龙门速度与计划中的预期龙门速度。跨四个直线加速器进行的测量通常表明,在恒定剂量率区域内,测量与计划之间的一致性达到1.5%以内,在计划的0.1 s之内进行剂量率调制。瞬时龙门速度测得在计划的(1 SD)的0.11°/ s之内。检测到一个直线加速器的错误,因为标称龙门速度未正确校准。此测试提供了一种实用的方法,可确保VMAT交付的关键方面的质量,包括剂量与机架角度以及机架速度控制。该方法可以用能够获取时间分辨的剂量信息的任何检测器来执行,该时间分辨的剂量信息可以与机架角度的测量同步。该测试实现了荷兰放射剂量学委员会(NCS)最新报告24的若干目标,该报告为全面VMAT质量保证提供了建议.PACS编号:87.55.Qr

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号