Comparison of four commercial devices for RapidArc and sliding window IMRT QA

机译:RapidArc和滑动窗口IMRT QA的四种商用设备的比较

获取原文

摘要

For intensity‐modulated radiation therapy, evaluation of the measured dose against the treatment planning calculated dose is essential in the context of patient‐specific quality assurance. The complexity of volumetric arc radiotherapy delivery attributed to its dynamic and synchronization nature require new methods and potentially new tools for the quality assurance of such techniques. In the present study, we evaluated and compared the dosimetric performance of EDR2 film and three other commercially available quality assurance devices: IBA I'MatriXX array, PTW Seven29 array and the Delta 4 array. The evaluation of these dosimetric systems was performed for RapidArc and IMRT deliveries using a Varian NovalisTX linear accelerator. The plans were generated using the Varian Eclipse treatment planning system. Our results showed that all four QA techniques yield equivalent results. All patient QAs passed our institutional clinical criteria of gamma index based on a 3% dose difference and 3 mm distance to agreement. In addition, the Bland‐Altman analysis was performed which showed that all the calculated gamma values of all three QA devices were within 5% from those of the film. The results showed that the four QA systems used in this patient‐specific IMRT QA analysis are equivalent. We concluded that the dosimetric systems under investigation can be used interchangeably for routine patient specific QA. PACS numbers: 87.55.Qr, 87.56.Fc
机译:对于强度调制放射疗法,在特定于患者的质量保证范围内,根据治疗计划的计算剂量评估测量剂量至关重要。由于其动态和同步特性,弧形放射治疗的复杂性需要新的方法和潜在的新工具来保证此类技术的质量。在本研究中,我们评估并比较了EDR2膜和其他三种商用质量保证设备的剂量学性能:IBA I'MatriXX阵列,PTW Seven29阵列和Delta 4阵列。这些剂量系统的评估是使用Varian NovalisTX线性加速器对RapidArc和IMRT交付进行的。计划是使用Varian Eclipse处理计划系统生成的。我们的结果表明,所有四种质量保证技术均能产生等效的结果。所有患者QA均通过了3%的剂量差和3 mm的一致距离,通过了我们的伽马指数机构临床标准。此外,进行了布兰德-奥尔特曼分析,结果表明,所有三个QA器件的所有伽马值计算值均与胶卷的5%之差。结果表明,此针对患者的IMRT QA分析中使用的四个QA系统是等效的。我们得出的结论是,所研究的剂量系统可以互换使用,以进行常规的患者特定QA。 PACS编号:87.55.Qr,87.56.Fc

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号