首页> 外文期刊>Journal of Analytical & Bioanalytical Techniques >Interference Effects of Excess Lithium on the Emission Signal of the Calcium and Strontium Ionic Lines during Inductively Coupled Plasma Emission Spectrometry: Simulation in Terms of the Simplified Collisional Radiative Recombination Rate Model
【24h】

Interference Effects of Excess Lithium on the Emission Signal of the Calcium and Strontium Ionic Lines during Inductively Coupled Plasma Emission Spectrometry: Simulation in Terms of the Simplified Collisional Radiative Recombination Rate Model

机译:电感耦合等离子体发射光谱法中过量锂对钙锶离子线发射信号的干扰效应:基于简化的碰撞辐射复合速率模型的模拟

获取原文
           

摘要

ICP-AES Ca and Sr ion line signals were measured on 0-30 μg/mL Ca and Sr solutions, in the absence and presence of 1000 μg/mL Li as easily ionizable interferent, and the effect of the interferent on analyte calibration curve simulated using a simplified collisional radiative recombination rate model. Close agreement between experiment and theory was obtained for the full range of concentrations studied in the case of SrII, while the theoretical Ca ion line calibration curve exhibited close agreement below 10 μg/mL Ca concentration in the test solution, and a positive deviation of up to 18% from the experimental curve at higher Ca concentrations. The data obtained successfully demonstrates the potential of the simplified collisional radiative recombination rate model for simulating the effects of easily ionizable interferents on high ionization potential analytes.
机译:在不存在和存在1000μg/ mL易于电离的干扰物Li的情况下,在0-30μg/ mL Ca和Sr溶液中测量ICP-AES Ca和Sr离子线信号,并模拟干扰物对分析物校准曲线的影响使用简化的碰撞辐射复合率模型。对于SrII,在所研究的整个浓度范围内,实验与理论之间都具有密切的一致性,而理论Ca离子线校准曲线在测试溶液中的Ca浓度低于10μg/ mL时显示出紧密的一致性,并且正偏差最大。在较高的Ca浓度下,从实验曲线可降低到18%。成功获得的数据证明了简化的碰撞辐射复合率模型在模拟容易电离的干扰物对高电离势分析物的影响方面的潜力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号