BackgroundA pulse oximeter is one of the most important monitors to save patients undergoing anesthesia and monitored sedation. The authors report a case of orthopedic surgery, in which interference of pulse oximetry occurred when using a Stealth Station? navigation system (Medtronic Sofamor Danek, Memphis, TN). Applying a black plastic shield (Masimo Ambient Shield: Masimo Corporation, Irvine, CA) completely eliminated the interference. Case presentationA 37-year-old male patient with a giant cell tumor of the left femur was scheduled to undergo curettage of the femur using an intraoperative CT three-dimensional imaging system (O-arm?) and Stealth Station? navigation system. During the surgery, the SpO2 value, which was maintained between 97 and 99% until the time, disappeared suddenly with abnormal pulse wave. Because a distortion in the SpO2 value was reproduced by repeated movement of cameras on the head of the Stealth Station? navigation system, we recognized that the interference signal was coming from the navigation system. To eliminate the infrared light, the pulse oximetry probe was covered with a black plastic shield and the interference was completely eliminated. ConclusionsThe Stealth Station? navigation system was found to interfere with the SpO2 value, and a black plastic shield was useful for eliminating the interfering signal. Anesthesiologists should understand the risk of interference by the neuro-navigation system and know how to solve the problem.
机译:背景脉搏血氧仪是最重要的监护仪之一,可以节省接受麻醉和镇静的患者。作者报告了一例骨科手术,其中使用隐形站时发生了脉搏血氧饱和度的干扰?导航系统(田纳西州孟菲斯的Medtronic Sofamor Danek)。使用黑色塑料屏蔽层(Masimo Ambient Shield:Masimo Corporation,Irvine,CA)完全消除了干扰。病例介绍一名37岁男性左股骨巨细胞瘤的患者计划使用术中CT三维成像系统(O型臂)和Stealth Station?进行刮除股骨。导航系统。手术期间,SpO 2 值一直保持在97%到99%之间,直到出现异常脉冲波时才突然消失。因为通过重复在隐形站头上移动摄像机来再现SpO 2 值的失真?导航系统,我们认识到干扰信号来自导航系统。为了消除红外光,脉搏血氧饱和度探头用黑色塑料屏蔽罩覆盖,并且完全消除了干扰。结论隐形站?导航系统被发现会干扰SpO 2 值,黑色塑料屏蔽层对于消除干扰信号很有用。麻醉师应了解神经导航系统干扰的风险,并知道如何解决该问题。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号