The Stealth StationTM Image Guidance System may interfere with pulse oximetry: (Le systeme de guidage par imagerie Stealth StationTM peut nuire a la sphygmo-oxymetrie).

机译:Stealth StationTM图像引导系统可能会干扰脉搏血氧饱和度:(Stealth StationTM图像引导系统可能会干扰脉搏血氧饱和度)。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: Interference on pulse oximetry can come from many sources. We found an additional source of interference from the Stealth Stationtrade mark. This article gives an overview of sources of pulse oximeter interference so that clinicians can better prevent them.Technical features: This article discusses the infrared disturbances caused by the Stealth Stationtrade mark. The Stealth Stationtrade mark is a frameless stereotactic positioning system that utilizes a three dimensional location system to measure the position of the patient and the surgical tools, and to relate those positions to previously recorded imaging. To understand the disturbance caused by the Stealth Stationtrade mark, we discuss its operation and that of pulse oximeter monitors. Pulse oximeter interference can come from volume artifacts, electrical and light noise, and can be caused by issues related to the patient. Because the passive Stealth Stationtrade mark contains a strong infrared light source, interference caused by light is a likely reason for the interference we noted. Pulse oximeters rely on the time-variant light signal modulated by arterial volume variations in the finger. Although relatively immune to static light sources, pulse oximeters are extremely sensitive to time-varying light sources. The light emitted by the passive Stealth Stationtrade mark is time-varying at 4 Hz and this is causing the pulse oximeter to provide invalid results. Shielding can generally be used to stop the light from the Stealth Stationtrade mark from being picked up by the pulse oximeter sensor. CONCLUSION: Infrared light interference can be very common, but is easily dealt with if one is aware of it.
机译:目的:对脉搏血氧饱和度的干扰可能来自许多来源。我们发现了隐身站商标的其他干扰源。本文概述了脉搏血氧仪的干扰源,以便临床医生可以更好地预防它们。技术特点:本文讨论了由隐身站商标引起的红外干扰。 Stealth Stationtrade商标是一种无框立体定位系统,该系统利用三维定位系统来测量患者和手术工具的位置,并将这些位置与先前记录的成像相关联。为了了解Stealth Station商标引起的干扰,我们讨论了它的操作以及脉搏血氧仪监测仪的操作。脉搏血氧饱和度干扰可能来自体积伪影,电噪声和光噪声,并且可能是由与患者有关的问题引起的。因为无源Stealth Station商标包含强红外光源,所以由光引起的干扰可能是我们注意到的干扰原因。脉搏血氧仪依赖于手指动脉体积变化调制的时变光信号。尽管脉搏血氧仪对静态光源相对免疫,但对时变光源极为敏感。无源Stealth Station商标发出的光在4 Hz时变,这使脉搏血氧饱和度计提供无效的结果。通常可以使用屏蔽来阻止来自隐身站商标的光被脉搏血氧仪传感器拾取。结论:红外光干扰可能非常普遍,但是如果人们意识到这一点很容易解决。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号