Evaluation of criteria of manual blood smear review following automated complete blood counts in a large university hospital

机译:大型大学医院自动全血细胞计数后的手动血液涂片检查标准评估

获取原文
   

获取外文期刊封面封底 >>

       

摘要

BackgroundThere is great interest in reducing the number of automated complete blood counts requiring manual blood smear reviews without sacrificing the quality of patient care. This study was aimed at evaluating and establishing appropriate screening criteria for manual blood smear reviews to improve the performance in a hematology laboratory.MethodA total of 1977 consecutive samples from the daily workload were used to evaluate four sets of screening criteria for manual blood smear reviews to identify samples with positive smear findings. Three sets of screening criteria were arbitrarily proposed in this study: Group 1 (narrow ranges), Group 2 (intermediate ranges), and Group 3 (wide limits) and one set (Group 4) was adapted from the International Society for Laboratory Hematology. All samples were run on Sysmex hematology analyzers and were investigated using manual blood smear reviews. Diagnostic accuracy and agreement were performed for each set of screening criteria, including an investigation of microscopic review rate and efficiency.ResultsThe microscopic review rates for Groups 1, 2, 3 and 4 were 73.85%, 54.52%, 46.33% and 46.38%, respectively; the false-negative rates were 0.50%, 1.97%, 2.73% and 3.95%, respectively. The efficiency and negative predictive values of Group 3 were 73.04% and 4.91%, respectively.ConclusionsGroup 3 had the best relationship between safety (false-negative rate: ≤3%) and efficiency to estimate the limits of automation in performing complete blood counts. This study strengthens the importance of establishing screening criteria for manual blood smear reviews, which account for the different contexts in which hematological determinations are performed. Each laboratory should optimize the screening criteria for manual blood smear reviews in order to maximize their efficiency and safety.
机译:背景技术对于减少需要人工涂片检查的自动全血细胞计数的数量而又不牺牲患者护理质量的关注。这项研究旨在评估和建立适当的手动血液涂片检查筛查标准,以提高血液学实验室的性能。方法每天工作量的1977个连续样本用于评估四组手动血液涂片检查的筛查标准,以评估识别涂片检查结果呈阳性的样品。本研究中任意提出了三套筛查标准:第1组(狭窄范围),第2组(中等范围)和第3组(宽范围),其中一组(第4组)是从国际实验室血液学协会改编的。所有样品均在Sysmex血液分析仪上运行,并使用手动血液涂片检查进行了调查。对每组筛查标准进行诊断准确性和一致性检查,包括对微观检查率和效率的调查。结果第1、2、3和4组的微观检查率分别为73.85%,54.52%,46.33%和46.38%。 ;假阴性率分别为0.50%,1.97%,2.73%和3.95%。第3组的效率和阴性预测值分别为73.04%和4.91%。结论第3组在安全性(假阴性率:≤3%)和效率之间具有最佳的关系,以估计自动执行全血细胞计数的极限。这项研究加强了为人工血液涂片检查建立筛查标准的重要性,该检查标准解释了进行血液学测定的不同环境。每个实验室应优化人工涂片检查的筛查标准,以最大程度地提高其效率和安全性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号