Effect of bone quality and implant surgical technique on implant stability quotient (ISQ) value

机译:骨质量和植入物手术技术对植入物稳定性商(ISQ)值的影响

获取原文

摘要

PURPOSE This study investigated the influence of bone quality and surgical technique on the implant stability quotient (ISQ) value. In addition, the influence of interfacial bone quality, directly surrounding the implant fixture, on the resonance frequency of the structure was also evaluated by the finite element analysis. MATERIALS AND METHODS Two different types of bone (type 1 and type 2) were extracted and trimmed from pig rib bone. In each type of bone, the same implants were installed in three different ways: (1) Compaction, (2) Self-tapping, and (3) Tapping. The ISQ value was measured and analyzed to evaluate the influence of bone quality and surgical technique on the implant primary stability. For finite element analysis, a three dimensional implant fixture-bone structure was designed and the fundamental resonance frequency of the structure was measured with three different density of interfacial bone surrounding the implant fixture. RESULTS In each group, the ISQ values were higher in type 1 bone than those in type 2 bone. Among three different insertion methods, the Tapping group showed the lowest ISQ value in both type 1 and type 2 bones. In both bone types, the Compaction groups showed slightly higher mean ISQ values than the Self-tapping groups, but the differences were not statistically significant. Increased interfacial bone density raised the resonance frequency value in the finite element analysis. CONCLUSION Both bone quality and surgical technique have influence on the implant primary stability, and resonance frequency has a positive relation with the density of implant fixture-surrounding bone.
机译:目的本研究调查了骨质量和手术技术对植入物稳定性商(ISQ)值的影响。此外,还通过有限元分析来评估直接围绕植入物固定装置的界面骨质量对结构共振频率的影响。材料与方法从猪肋骨中提取并修剪两种不同类型的骨(1型和2型)。在每种类型的骨头中,以三种不同的方式安装相同的植入物:(1)压紧,(2)自攻和(3)自攻。测量并分析ISQ值,以评估骨质量和手术技术对植入物基本稳定性的影响。为了进行有限元分析,设计了三维植入物固定器-骨结构,并通过围绕植入物固定器的三种不同密度的界面骨来测量该结构的基本共振频率。结果在每组中,ISQ值在1型骨中均高于2型骨。在三种不同的插入方法中,“攻牙”组在1型和2型骨骼中均显示出最低的ISQ值。在两种骨类型中,压实组的平均ISQ值均高于自攻组,但差异无统计学意义。界面骨密度的增加在有限元分析中提高了共振频率值。结论骨骼质量和手术技术都对植入物的初步稳定性有影响,共振频率与植入物周围的骨密度呈正相关。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号