Inappropriate Implantable Cardioverter-Defibrillator Shocks Attributed to Alternating-Current Leak in a Swimming Pool

机译:游泳池交流电泄漏引起的不适当的植入式心脏复律除颤器电击

获取原文
           

摘要

Implantable cardioverter-defibrillators (ICDs) are the standard of care for preventing sudden cardiac death in patients who are predisposed to malignant ventricular arrhythmias. Causes of inappropriate ICD shock include equipment malfunction, improper arrhythmia evaluation, misinterpretation of myopotentials, and electromagnetic interference. As the number of implanted ICDs has increased, other contributors to inappropriate therapy have become known, such as minimal electrical current leaks that mimic ventricular fibrillation. We present the case of a 63-year-old man with a biventricular ICD who received 2 inappropriate shocks, probably attributable to alternating-current leaks in a swimming pool. In addition, we discuss ICD sensitivity and offer recommendations to avoid similar occurrences.
机译:植入式心脏复律除颤器(ICD)是预防易患恶性室性心律不齐的患者突发心源性死亡的护理标准。 ICD不当电击的原因包括设备故障,心律失常评估不当,肌电势误解以及电磁干扰。随着植入的ICD数量的增加,其他导致不当治疗的因素也已广为人知,例如模仿心室纤颤的最小电流泄漏。我们介绍了一个患有双室ICD的63岁男子的案例,该人受到2次不适当的电击,可能是由于游泳池中的交流电泄漏造成的。此外,我们讨论了ICD敏感性,并提供了避免类似情况的建议。