【24h】

Inappropriate implantable cardioverter-defibrillator shock due to external alternating current leak: report of two cases.

机译:由于外部交流泄漏引起的植入式心脏复律除颤器不当电击:两例报道。

获取原文
获取原文并翻译 | 示例
           

摘要

Two cases are reported of inappropriate implantable cardioverter-defibrillator shocks due to external alternating current leak. Electromagnetic interference (EMI) can mimic cardiac signals and cause inappropriate implantable cardioverter-defibrillator (ICD) shocks. EMI can arise from the normal functioning of electrical appliances or from alternating current leak. The two cases had inappropriate ICD shocks due to alternating current leak from a power drill in one and a washing machine in the other. The need for detailed advice on handling electrical equipment is emphasized.
机译:据报道有两例由于外部交流电泄漏而引起的植入式心脏复律除颤器不当电击。电磁干扰(EMI)会模仿心脏信号,并引起不适当的植入式心脏复律除颤器(ICD)电击。 EMI可能是由于电器的正常运行或交流电泄漏引起的。由于一台电钻和另一台洗衣机的交流电泄漏,这两种情况都有不适当的ICD冲击。强调需要有关处理电气设备的详细建议。