Use of 100 kV versus 120 kV in computed tomography pulmonary angiography in the detection of pulmonary embolism: effect on radiation dose and image quality

机译:在计算机断层扫描肺血管造影中使用100 kV和120 kV对肺栓塞的检测:对放射剂量和图像质量的影响

获取原文
           

摘要

>>Objective: To determine the effective radiation dose and image quality resulting from 100 versus 120 kilovoltage (kV) protocols among patients referred for computed tomography pulmonary angiography (CTPA).>>Methods: Sixty-six patients with clinical suspicion of pulmonary embolism (PE) were prospectively enrolled. Two CTPA protocols (group A: n=33, 100 kV/115 mAs; group B: n=33, 120 kV/90 mAs) were compared. Two experienced radiologists assessed image quality in terms of diagnostic performance and effect of artefacts. Image quality parameters [CT attenuation, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR)] and effective radiation dose between the two protocols were compared.>>Results: The contrast enhancement in central and peripheral pulmonary arteries was significantly higher in group A than in group B (P0.001) with the identical SNR (P=0.26), whereas the CNR was significantly higher in group A than in group B (P0.001). The effective radiation dose for the 100 and 120 kV scans was 3.2 and 6.8 mSv, respectively.>>Conclusions: Reducing the tube voltage from 120 to 100 kV in CTPA allows a significant reduction of radiation dose without significant loss of diagnostic image quality.
机译:> >目标:确定在计算机断层扫描肺血管造影(CTPA)患者中100 kV与120 kV(kV)方案产生的有效辐射剂量和图像质量。 > >方法:前瞻性纳入了66例临床怀疑肺栓塞(PE)的患者。比较了两个CTPA协议(A组:n = 33,100 kV / 115 mAs; B组:n = 33,120 kV / 90 mAs)。两名经验丰富的放射科医生根据诊断性能和伪影效果评估了图像质量。比较了两种方案之间的图像质量参数[CT衰减,信噪比(SNR)和对比噪声比(CNR)]和有效辐射剂量。 > >结果: 在具有相同SNR(P = 0.26)的情况下,A组中央和外周肺动脉的对比度增强显着高于B组(P <0.001),而A组的CNR显着高于B组B(P <0.001)。 100和120 kV扫描的有效辐射剂量分别为3.2和6.8 mSv。 > >结论:将CTPA中的管电压从120 kV降低到100 kV可以显着降低辐射剂量,而不会显着降低诊断图像质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号