首页> 外文期刊>Practical Laboratory Medicine >Have you stress tested your assay?
【24h】

Have you stress tested your assay?

机译:您压力测试了吗?

获取原文
       

摘要

Objectives When a clinical assay is stressed with extraordinarily high volume of specimens over a short period of time, extra caution may be needed to avoid systematic errors and biases. Here we report our experience with a HgbA1c assay used for high volume wellness screening purpose, to illustrate the importance of stress testing during assay validation. Design and Methods Over 15,000 whole blood specimens were tested for HgbA1c in a period of 2 months. HgbA1c was tested by an immunoturbidimetric method on a high through-put automation line. The HgbA1c population distribution in our study was compared to that from the NHANES database. Daily distributions of HgbA1c values ≥6%, means and medians were plotted. Correlation studies were performed between the high through-put immunoturbidimetric assay and a medium through-put HPLC method. Results We observed a shift of HgbA1c distribution to the higher values compared to the NHANES. A bias of 15–20% was noted from further stress testing where large number of samples were batched and tested using the immunoturbidimetric assay. A 5–7% higher bias remained after implementing a cuvette washing program after each HgbA1c sample. We hypothesized this bias was caused by build-up of blood cell fragments in the cuvettes when continuous whole blood samples are run through the system. Our experience suggests stress testing needs to be incorporated early in the test validation process for high volume batched screening applications. This seemingly extra validation step may save significant troubleshooting and retesting efforts down the road.
机译:目的当短期内用大量标本进行临床分析时,可能需要格外小心,以免出现系统误差和偏差。在这里,我们报告用于大型健康筛查目的的HgbA1c测定的经验,以说明测定验证期间进行压力测试的重要性。设计与方法在2个月的时间内,测试了超过15,000个全血标本中的HgbA1c。 HgbA1c在高通量自动化生产线上通过免疫比浊法进行了测试。我们研究中的HgbA1c人群分布与NHANES数据库中的分布进行了比较。 HgbA1c值的日分布≥6%,绘制平均值和中位数。在高通量免疫比浊法和中通量HPLC方法之间进行了相关研究。结果我们观察到与NHANES相比,HgbA1c分布向更高的值移动。进一步的压力测试表明有15-20%的偏差,其中大量样品被分批并使用免疫比浊法进行了测试。每个HgbA1c样品实施比色皿清洗程序后,偏差仍会增加5–7%。我们假设这种偏见是由连续的全血样本流经系统时比色皿中血细胞碎片的堆积引起的。我们的经验表明,对于大批量分批筛选应用,需要在测试验证过程的早期就进行压力测试。这个看似多余的验证步骤可以节省大量的故障排除和重新测试工作。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号