首页> 外文期刊>Results in Physics >Application of a dual energy X-ray imaging method on breast specimen
【24h】

Application of a dual energy X-ray imaging method on breast specimen

机译:双能X线成像技术在乳腺标本上的应用

获取原文
           

摘要

Highlights ? Dual energy images were obtained from a formalin-fixed breast cancer specimen. ? Dual energy images were compared to the routine digital mammographic image. ? Increased calcification information was identified using the dual energy method. Abstract The purpose of this study was to evaluate a dual energy method, developed by our group, on a breast cancer specimen. A modified radiographic X-ray tube combined with a high resolution complementary metal-oxide semiconductor (CMOS) active pixel sensor (APS) X-ray detector was used. A 40/70kV spectral combination was filtered with 100μm cadmium (Cd) and 1000μm copper (Cu) for the low/high-energy combination. Dual energy images were obtained from a formalin-fixed breast cancer specimen for various entrance surface doses (ESD). Initial results showed that the DE images were directly comparable with the mammographic image and similar or even increased calcification information was identified, with mean glandular dose values at acceptable levels.
机译:强调 ?从福尔马林固定的乳腺癌标本中获得双能图像。 ?将双能图像与常规数字乳腺摄影图像进行比较。 ?使用双重能量方法确定了增加的钙化信息。摘要这项研究的目的是评估我们小组在乳腺癌样本上开发的双重能量方法。使用改进的射线X射线管与高分辨率互补金属氧化物半导体(CMOS)有源像素传感器(APS)X射线检测器相结合。将40 / 70kV光谱组合用100μm镉(Cd)和1000μm铜(Cu)过滤以实现低/高能组合。从福尔马林固定的乳腺癌标本中获得各种能量入射表面剂量(ESD)的双能图像。初步结果表明,DE图像可直接与乳腺X线图像相媲美,并且可以确定相似或什至增加的钙化信息,平均腺体剂量值可以接受。