首页> 外文期刊>Revista Brasileira de Hematologia e Hemoterapia >Assessment of blood sample stability for complete blood count using the Sysmex XN-9000 and Mindray BC-6800 analyzers
【24h】

Assessment of blood sample stability for complete blood count using the Sysmex XN-9000 and Mindray BC-6800 analyzers

机译:使用Sysmex XN-9000和Mindray BC-6800分析仪评估全血样的血样稳定性

获取原文
       

摘要

BACKGROUND: Different hematological analyzers have different analytical performances that are often reflected in the criteria for sample stability of the complete blood count. This study aimed to assess the stability of several hematological parameters using the XN-9000 Sysmex and BC-6800 Mindray analyzers. METHODS: The impact of storage at room temperature and 4 ?°C was evaluated after 2, 4, 6, 8, 24, 36 and 48 h using ten normal and 40 abnormal blood samples. The variation from the baseline measurement was evaluated by the Steel-Dwass-Critchlow-Fligner test and by Bland-Altman plots, using quality specifications and critical difference as the total allowable variation. RESULTS: Red blood cells and reticulocyte parameters (i.e. hematocrit, mean corpuscular volume, mean corpuscular hemoglobin concentration, red blood cell distribution width, immature reticulocyte fractions, low-fluorescence reticulocytes, middle-fluorescence reticulocytes, high fluorescence mononuclear cells) showed less stability compared to leukocyte and platelet parameters (except for monocyte count and mean platelet volume). The bias for hematocrit, mean corpuscular volume, mean corpuscular hemoglobin concentration and red blood cell distribution width coefficient of variation was higher than the critical difference after 8 h using both analyzers. CONCLUSION: Blood samples measured with both analyzers do not show analytically significant changes in up to 2 h of storage at room temperature and 4 ?°C. However, the maximum time for analysis can be extended for up to 8 h when the bias is compared to the critical difference.
机译:背景:不同的血液分析仪具有不同的分析性能,这些性能通常反映在全血细胞计数的样品稳定性标准中。这项研究的目的是使用XN-9000 Sysmex和BC-6800 Mindray分析仪评估几种血液学参数的稳定性。方法:使用10份正常和40份异常血样,在2、4、6、8、24、36和48小时后评估室温和4°C下储存的影响。使用质量规格和临界差作为总允许变化,通过Steel-Dwass-Critchlow-Fligner测试和Bland-Altman图评估了基线测量的变化。结果:红细胞和网织红细胞参数(即血细胞比容,平均红细胞体积,平均红细胞血红蛋白浓度,红细胞分布宽度,未成熟网织红细胞分数,低荧光网织红细胞,中等荧光网织红细胞,高荧光单核细胞)显示出稳定性较差白细胞和血小板参数(单核细胞计数和平均血小板体积除外)。使用这两种分析仪后,血细胞比容,平均红细胞体积,平均红细胞血红蛋白浓度和红细胞分布宽度变异系数的偏差均高于8小时后的临界差。结论:使用两种分析仪测量的血液样本在室温和4°C的长达2小时的存储中均未显示出分析上的显着变化。但是,将偏差与临界差值进行比较时,最长分析时间最多可以延长8小时。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号