首页> 外文期刊>Radiation Oncology Journal >LiF TLD in TLD Holder for In Vivo Dosimetry
【24h】

LiF TLD in TLD Holder for In Vivo Dosimetry

机译:TLD支架中的LiF TLD用于体内剂量测定

获取原文
           

摘要

PURPOSE: LiF TLD has a problem to be used in vivo dosimetry because of the toxic property of LiF. The aim of this study is to develop new dosimeter with LiF TLD to be used in vivo dosimetry. MATERIALS AND METHODS: We designed and manufactured the teflon box(here after TLD holder) to put TLD in. The external size of TLD holder is 4x4x1 mm3. To estimate the effect of TLD holder on TLD response for radiation, the linearity of TLD response to nominal dose were measured for TLD in TLD holder. Measurement were performed in the 10 MV x-ray beam with LiF TLD using a solid water phantom at SSD of 100 cm. Percent Depth Dose (PDD) and Tissue-Maximum Ratio (TMR) with varying phantom thickness on TLD were measured to find the effect of TLD holder on the dose coefficient used for dose calculation in radiation therapy. RESULTS: The linearity of response of TLD in TLD holder to the nominal dose was improved than TLD only used as dosimeter. And in various measurement conditions, it makes a marginnal difference between TLD in TLD holder and TLD only in their responses. CONCLUSION: It was proven that the TLD in TLD holder as a new dosimetry could be used in vivo dosimetry.
机译:目的:由于LiF的毒性,LiF TLD在体内剂量测定中存在问题。这项研究的目的是开发具有LiF TLD的新型剂量计,用于体内剂量测定。材料和方法:我们设计并制造了用于放置TLD的特氟龙盒(在此之后为TLD支架)。TLD支架的外部尺寸为4x4x1 mm3。为了估算TLD支架对TLD辐射响应的影响,测量了TLD支架中TLD相对于标称剂量的TLD响应的线性。使用LiF TLD在10 MV X射线束中使用固态水体模在100 cm的SSD上进行测量。测量了不同的幻影厚度对DLD的百分比深度剂量(PDD)和组织最大比(TMR),以发现TLD支架对用于放射治疗剂量计算的剂量系数的影响。结果:与仅用作剂量计的TLD相比,TLD持有人中TLD对标称剂量的响应线性有所改善。并且在各种测量条件下,仅在TLD持有人的TLD和TLD的响应之间产生边际差异。结论:已证明TLD支架中的TLD作为一种新的剂量测定方法可用于体内剂量测定。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号