Correlation between target motion and the dosimetric variance of breast and organ at risk during whole breast radiotherapy using 4DCT

机译:使用4DCT进行全乳放疗期间目标运动与有风险的乳房和器官的剂量学差异之间的相关性

获取原文

摘要

Purpose The purpose of this study was to explore the correlation between the respiration-induced target motion and volume variation with the dosimetric variance on breast and organ at risk (OAR) during free breathing. Methods and materials After breast-conserving surgery, seventeen patients underwent respiration-synchronized 4DCT simulation scans during free breathing. Treatment planning was constructed using the end inspiration scan, then copied and applied to the other phases and the dose distribution was calculated separately to evaluate the dose-volume histograms (DVH) parameters for the planning target volume (PTV), ipsilateral lung and heart. Results During free breathing, the treated breast motion vector was 2.09?±?0.74 mm, and the volume variation was 3.05?±?0.94%. There was no correlation between the breast volume and target/OAR dosimetric variation (|r|?=?0.39?~?0.48). In the anteroposterior, superoinferior and vector directions, breast movement correlated well with the mean PTV dose, conformal index, and the lung volume receiving high dose (|r|?=?0.651-0.975); in the superoinferior and vector directions, breast displacement only correlated with the heart volume receiving >5 Gy (V5) (r?=??0.795, 0.687). The lung volume and the lung volume receiving high dose correlated reasonably well (r?=?0.655?~?0.882), and a correlation only existed between heart volume and V5 (r?=??0.701). Conclusion Target movement correlated well with the target/OAR dosimetric variation in certain directions, indicating that whole breast IMRT assisted by breathing control or respiratory-adapted gated treatment promotes the accuracy of dose delivery during radiotherapy. During free breathing, the effect of breast volume variation can be ignored in whole breast IMRT.
机译:目的这项研究的目的是探讨呼吸引起的目标运动和体积变化与自由呼吸期间乳房和高危器官(OAR)的剂量学差异之间的相关性。方法和材料保乳手术后,在自由呼吸期间对17位患者进行了呼吸同步的4DCT模拟扫描。使用最终吸气扫描构建治疗计划,然后复制并应用于其他阶段,并分别计算剂量分布,以评估计划目标体积(PTV),同侧肺和心脏的剂量体积直方图(DVH)参数。结果在自由呼吸过程中,治疗后的乳房运动矢量为2.09±±0.74mm,体积变化为3.05±±0.94%。乳房体积与目标/ OAR剂量变化之间没有相关性(| r |?=?0.39?〜?0.48)。在前后,上,下方向和矢量方向上,乳房运动与平均PTV剂量,保形指数以及接受高剂量的肺容积密切相关(| r |?=?0.651-0.975);在上下方向和矢量方向上,乳房移位仅与接收> 5 Gy(V5)的心脏体积相关(r?=?0.795,0.687)。肺容积与接受大剂量的肺容积相关性良好(r≥0.655≥0.882),并且仅在心脏体积与V5之间存在相关(r≥0.701)。结论靶标运动与靶标/ OAR剂量在某些方向上的变化具有良好的相关性,表明通过呼吸控制或适应呼吸的门控治疗辅助全乳IMRT可以提高放疗期间剂量的准确性。自由呼吸期间,整个乳房IMRT中乳房体积变化的影响可以忽略。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号