首页> 外文期刊>Revista Brasileira de Anestesiologia >Falha no funcionamento do ressuscitador manual autoinflável devido à presen?a de secre??es pulmonares ressecadas
【24h】

Falha no funcionamento do ressuscitador manual autoinflável devido à presen?a de secre??es pulmonares ressecadas

机译:由于存在切除的肺部分泌物,自动充气人工复苏器无法正常工作

获取原文
           

摘要

BACKGROUND AND OBJECTIVES: Self-inflating manual resuscitators (SIMR) are devices used to ventilate patients with ventilatory needs. These devices consist of a set of valves that work sequentially, and changes in their function may be harmful to patients. CASE REPORT: During the use of SIMR, it was observed that the valve remained fixed due to the presence of dried pulmonary secretion, which made it impossible to ventilate the patient, but not manipulate the compressible unit. CONCLUSIONS: This situation reaffirms that the resuscitator is a device that should be used by trained professionals because, although changes in functioning are rare, they can be potentially fatal.
机译:背景和目的:自充气手动复苏器(SIMR)是用于为有呼吸需要的患者通气的设备。这些设备由一组按顺序工作的阀组成,其功能变化可能对患者有害。病例报告:在使用SIMR期间,观察到由于干燥的肺部分泌物导致瓣膜保持固定,这使患者无法通气,但无法操纵可压缩单元。结论:这种情况重申了复苏器是一种应由受过训练的专业人员使用的装置,因为尽管功能上的改变很少见,但它们可能是致命的。