首页> 外文期刊>Revista Brasileira de Anestesiologia >Compara??o da FiO2 fornecida por sete modelos de sistema bal?o-máscara auto-inflável
【24h】

Compara??o da FiO2 fornecida por sete modelos de sistema bal?o-máscara auto-inflável

机译:七个自充气式气球罩系统提供的FiO2的比较

获取原文
           

摘要

BACKGROUND AND OBJECTIVES: Since resuscitators with self-inflating bag-mask systems manufactured and/or commercialized in Brazil are widely available and used in health services, both out- and intra-hospitals, the objective of this study was to determine the O2 fractions delivered by seven resuscitators receiving different O2 flows. METHODS: Seven resuscitators with self-inflating bag-mask systems were tested at the Respiratory Unit of the HC/UNICAMP. A wall O2 flowmeter was connected to the resuscitator that received an O2 flow of 1, 5, 10, and 15 L.min-1 and those were connected to a test lung. Resuscitators capable of being connected to an O2 reservoir were tested with and without this accessory. Twenty consecutive measurements were performed and the mean determined. RESULTS: Only one resuscitator delivered and O2 fraction slightly below the accepted limit (0.80) when used with the O2 reservoir. Without this device, all resuscitators achieved the minimal limit of O2 fraction (0.40). Resuscitators not capable of being connected to an O2 reservoir delivered a higher O2. CONCLUSIONS: All resuscitators capable of being connected to an O2 reservoir delivered a higher O2 concentration when connected to this device. Resuscitators that do not have this capability delivered a higher O2 concentration than the ones that could be connected to this device but are used without it.
机译:背景和目的:由于在巴西制造和/或商业化的带有自充气袋式口罩系统的复苏器在医院外和医院内都可广泛获得并用于医疗服务,因此本研究的目的是确定所输送的O2分数由七个接受不同氧气流量的复苏器。方法:在HC / UNICAMP的呼吸科测试了七台带有自充气袋口罩系统的复苏器。将壁式氧气流量计连接到复苏器,复苏器接收的氧气流量为1、5、10和15 L.min-1,而复苏器则连接到测试肺。在有或没有此附件的情况下,都测试了能够连接到氧气储罐的复苏器。进行了二十次连续测量并确定了平均值。结果:与O2储罐一起使用时,仅输送了一个复苏器,O2分数略低于可接受的极限(0.80)。没有此设备,所有复苏器的氧气含量最低限度为(0.40)。无法连接到O2储罐的复苏器释放了更高的O2。结论:所有能够连接到氧气储存器的复苏器在连接到该设备时都可以提供更高的氧气浓度。没有此功能的复苏器提供的O2浓度要高于可以连接到该设备但没有使用的复苏器。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号