Prospective Comparison of Liver Stiffness Measurements between Two Point Shear Wave Elastography Methods: Virtual Touch Quantification and Elastography Point Quantification

机译:两点剪切波弹性成像方法之间的肝脏刚度测量的前瞻性比较:虚拟触摸定量和弹性成像点定量

获取原文
           

摘要

Objective To prospectively compare technical success rate and reliable measurements of virtual touch quantification (VTQ) elastography and elastography point quantification (ElastPQ), and to correlate liver stiffness (LS) measurements obtained by the two elastography techniques. Materials and Methods Our study included 85 patients, 80 of whom were previously diagnosed with chronic liver disease. The technical success rate and reliable measurements of the two kinds of point shear wave elastography (pSWE) techniques were compared by χ2 analysis. LS values measured using the two techniques were compared and correlated via Wilcoxon signed-rank test, Spearman correlation coefficient, and 95% Bland-Altman limit of agreement. The intraobserver reproducibility of ElastPQ was determined by 95% Bland-Altman limit of agreement and intraclass correlation coefficient (ICC). Results The two pSWE techniques showed similar technical success rate (98.8% for VTQ vs. 95.3% for ElastPQ, p = 0.823) and reliable LS measurements (95.3% for VTQ vs. 90.6% for ElastPQ, p = 0.509). The mean LS measurements obtained by VTQ (1.71 ± 0.47 m/s) and ElastPQ (1.66 ± 0.41 m/s) were not significantly different ( p = 0.209). The LS measurements obtained by the two techniques showed strong correlation ( r = 0.820); in addition, the 95% limit of agreement of the two methods was 27.5% of the mean. Finally, the ICC of repeat ElastPQ measurements was 0.991. Conclusion Virtual touch quantification and ElastPQ showed similar technical success rate and reliable measurements, with strongly correlated LS measurements. However, the two methods are not interchangeable due to the large limit of agreement.
机译:目的前瞻性比较虚拟触摸定量(VTQ)弹性成像和弹性成像点定量(ElastPQ)的技术成功率和可靠测量结果,并关联两种弹性成像技术获得的肝硬度(LS)测量值。材料和方法我们的研究包括85位患者,其中80位先前被诊断出患有慢性肝病。通过χ 2 分析比较了两种点剪切波弹性成像技术(pSWE)的技术成功率和可靠度。比较了通过两种技术测得的LS值,并通过Wilcoxon符号秩检验,Spearman相关系数和95%Bland-Altman协议极限进行了相关。 ElastPQ的观察者内部可重复性由95%的Bland-Altman一致性极限和组内相关系数(ICC)确定。结果两种pSWE技术显示出相似的技术成功率(VTQ为98.8%,ElastPQ为95.3%,p = 0.823)和可靠的LS测量(VTQ为95.3%,ElastPQ为90.6%,p = 0.509)。 VTQ(1.71±0.47 m / s)和ElastPQ(1.66±0.41 m / s)获得的平均LS测量值没有显着差异(p = 0.209)。两种技术获得的最小二乘法测量结果显示出很强的相关性(r = 0.820)。另外,两种方法的一致性的95%限制为平均值的27.5%。最终,重复ElastPQ测量的ICC为0.991。结论虚拟触摸定量和ElastPQ显示出相似的技术成功率和可靠的测量结果,并且LS测量值高度相关。但是,由于协议的限制,这两种方法不可互换。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号