Dual-Energy Computed Tomography Arthrography of the Shoulder Joint Using Virtual Monochromatic Spectral Imaging: Optimal Dose of Contrast Agent and Monochromatic Energy Level

机译:虚拟单色光谱成像的肩关节双能计算机断层摄影术:最佳剂量的造影剂和单色能级

获取原文
           

摘要

Objective To optimize the dose of contrast agent and the level of energy for dual-energy computed tomography (DECT) arthrography of the shoulder joint and to evaluate the benefits of the optimized imaging protocol. Materials and Methods Dual-energy scans with monochromatic spectral imaging mode and conventional single energy scans were performed on a shoulder phantom with 10 concentrations from 0 to 210 mg/mL of iodinated contrast medium at intervals of 15 or 30 mg/mL. Image noise, tissue contrast, and beam hardening artifacts were assessed to determine the optimum dose of contrast agent and the level of monochromatic energy for DECT shoulder arthrography in terms of the lowest image noise and the least beam hardening artifacts while good tissue contrast was maintained. Material decomposition (MD) imaging for bone-iodine differentiation was qualitatively assessed. The optimized protocol was applied and evaluated in 23 patients. Results The optimal contrast dose and energy level were determined by the phantom study at 60 mg/mL and 72 keV, respectively. This optimized protocol for human study reduced the image noise and the beam-hardening artifacts by 35.9% and 44.5%, respectively. Bone-iodine differentiation by MD imaging was not affected by the iodine concentration or level of energy. Conclusion Dual-energy scan with monochromatic spectral imaging mode results in reduced image noise and beam hardening artifacts.
机译:目的为肩关节的双能计算机断层扫描(DECT)关节造影术优化造影剂剂量和能量水平,并评估优化成像方案的益处。材料和方法用单色光谱成像模式进行双能量扫描和常规的单能量扫描是在15至30 mg / mL的间隔内,以10浓度的0至210 mg / mL碘化造影剂对肩部体模进行的。评估图像噪声,组织对比度和束硬化伪像,以确定最低的图像噪声和最少的束硬化伪像,同时保持良好的组织对比度,从而确定DECT肩关节造影术的最佳造影剂剂量和单色能量水平。对骨碘分化的材料分解(MD)成像进行了定性评估。应用优化方案对23例患者进行评估。结果通过体模研究分别确定了最佳对比剂量和能量水平,分别为60 mg / mL和72 keV。这种用于人类研究的优化协议分别将图像噪声和光束硬化伪像分别降低了35.9%和44.5%。通过MD成像进行的骨碘分化不受碘浓度或能量水平的影响。结论具有单色光谱成像模式的双能量扫描可减少图像噪声和光束硬化伪影。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号