首页> 外文期刊>Korean Circulation Journal >A Study on Real-Time Arrhythmia Detection System(II)(Part II. Signal Process Algorithm)
【24h】

A Study on Real-Time Arrhythmia Detection System(II)(Part II. Signal Process Algorithm)

机译:实时心律失常检测系统研究(二)(信号处理算法)

获取原文
           

摘要

We designed software system to detect intermittent arrhythmia from the patients automatically. To overcome the difficulties on real-time processing, we used four variavles; heart rate, QS width, morpholigy and axis. Templete generation routine was designed to produce patient's individual normal ECG pattern and store in the computer. The most important point of automatic detection is to discriminate the Q and J point. It is done by linear phase differentiation, the point that five point gradient is within a minimum value of 6Hz curve gradient is defined as Q and J points. The morphology calculation is simplified by simple addition and substraction, making real-time processing by modified cross correlation method. We find that both value of morphology and QS width will be used as a parameter of datermining the abnormalities of ECG signal automatically.
机译:我们设计了可以自动检测患者间歇性心律失常的软件系统。为了克服实时处理的困难,我们使用了四个变量。心率,QS宽度,形态和轴。 Templete生成例程旨在生成患者的个人正常ECG模式并存储在计算机中。自动检测最重要的一点是区分Q和J点。通过线性相位微分来完成,将五点梯度在6Hz曲线梯度的最小值内的点定义为Q点和J点。通过简单的加法和减法简化了形态计算,并通过改进的互相关方法进行了实时处理。我们发现形态学值和QS宽度都将被用作自动确定ECG信号异常的参数。