Evaluation of dental X-ray apparatus in terms of patient exposure to ionizing radiation

机译:根据患者对电离辐射的暴露程度评估牙科X射线设备

获取原文
           

摘要

Background: The use of X-ray in dental procedures causes exposure of the patient to ionizing radiation. This exposure dependsprimarily on the parameters used in tooth examination. The aim of the study was to determine the patients exposure and to assessthe technical condition of X-ray tubes. Material and Methods: Seventeen hundred dental offices were covered by the questionnairesurvey and 740 questionnaires were sent back. Direct measurements were performed in 100 units by using the thermoluminescentdetectors and X-ray films. Results: The results showed that the most commonly used exposure time is 0.22±0.16 s. Theaverage entrance dose for the parameters used most commonly by dentists is 1.7±1.4 mGy. The average efficiency of X-ray tubeestimated on the basis of exposures is 46.5±23.7 μGy/mAs. Conclusions: The study results indicate that the vast majority of X-raytubes meet the requirements specified in the binding regulations. Med Pr 2017;67(4):491–496
机译:背景:在牙科手术中使用X射线会使患者暴露于电离辐射中。这种暴露主要取决于牙齿检查中使用的参数。该研究的目的是确定患者的暴露情况并评估X射线管的技术状况。材料和方法:1.7万个牙科诊所接受了问卷调查,并发回了740份问卷。使用热发光检测器和X射线胶片以100个单位进行直接测量。结果:结果表明,最常用的曝光时间为0.22±0.16 s。牙医最常用的参数的平均进入剂量为1.7±1.4 mGy。根据曝光量估算的X射线管平均效率为46.5±23.7μGy/ mAs。结论:研究结果表明,绝大多数X射线管均符合约束法规中规定的要求。医学杂志2017; 67(4):491–496