Investigation of Exposure Factors for Various Breast Composition and Thicknesses in Digital Screening Mammography Related to Breast Dose

机译:与乳房剂量相关的数字化乳腺X线摄影中各种乳房组成和厚度的暴露因素调查

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Objective: To investigate the effect of exposure factors used in digital screening mammography on image quality of different breast compositions. Material and Methods: A digital mammography unit, with tungsten (W) as target, rhodium (Rh) and silver (Ag) as filters, and amorphous selenium detectors, was used to image Computerized Imaging Reference Systems (CIRS) Model 12A phantoms of thickness 4, 5, and 6 cm. Images of each phantom were obtained using target-filter combinations of W/Rh and W/Ag, at 28, 30, and 32 kVp. Images were evaluated by 5 senior technologists with experience in mammography. Image scores were assigned, for each type of feature present in the phantom. Statistical analysis was performed using nonparametric tests to compare sets of image scores at p = 0.05. Results: A small but statistically significant improvement was detected in the visibility of microcalcifications (8.8 ± 0.2; p = 0.031) for the W/Rh combination but this did not show any differences in the visibility of masses or fibers. The entrance skin dose (ESD) and mean glandular dose (MGD) were lower for the W/Ag (ESD = 1.30–3.70; MGD = 0.44–0.93 mGy) combination compared to W/Rh (ESD = 1.66–5.40; MGD = 0.52–1.12 mGy). The Mann-Whitney test revealed that 30-kV exposure with the W/Rh combination showed a significantly better visibility of specks in the 30/70 phantom compared to other exposures. Conclusion: The use of an Rh filter showed a better image quality for all phantoms. 28 and 30 kVp with the W/Rh combination provided a slightly better image quality, and the MGD is less than 1.2 mGy.
机译:目的:探讨在乳腺数字筛查中使用的暴露因素对不同乳腺成分的图像质量的影响。材料和方法:以钼(W)为靶标,铑(Rh)和银(Ag)为滤光片以及非晶态硒探测器的数字化乳腺X线摄影装置,对厚度为12A的计算机影像参考系统(CIRS)模型进行成像。 4、5和6厘米。使用W / Rh和W / Ag的目标滤波器组合以28、30和32 kVp获得每个体模的图像。图像由5位具有乳腺摄影经验的高级技术人员评估。为体模中存在的每种特征分配了图像分数。使用非参数测试进行统计分析,以比较p = 0.05时的图像分数集。结果:对于W / Rh组合,微钙化的可见度(8.8±0.2; p = 0.031)有微小但统计学上的显着改善,但这在质量或纤维的可见度上没有任何差异。 W / Ag(ESD = 1.30–3.70; MGD = 0.44–0.93 mGy)组合的入口皮肤剂量(ESD)和平均腺体剂量(MGD)低于W / Rh(ESD = 1.66–5.40; MGD = 0.52-1.12 mGy)。 Mann-Whitney测试表明,与其他暴露相比,W / Rh组合对30kV的暴露显示出30/70幻象中斑点的可见性明显更好。结论:Rh滤镜的使用为所有体模显示了更好的图像质量。 W / Rh组合使用28 kVp和30 kVp可以提供更好的图像质量,而MGD小于1.2 mGy。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号