摘要

Background:The objective of the study was to evaluate hearing threshold errors induced by the use of a new acoustical protocol (Cochlea-Scan).Material/Methods:Thirty-five normally hearing and 30 hearing-impaired adults were used in the study. Hearing threshold levels were assessed by pure tone audiometry (PTA) and the Cochlea-Scan (Fischer-Zoth/Natus) DPOAE protocols.Results:In the normal-hearing group, the mean difference in hearing levels at 2.0, 3.0, and 4.0 kHz was below 8 dB HL. At 1.5 kHz the mean difference was around 9 dB HL, while at 6.0 kHz the behavioral threshold was overestimated by 2 dB HL. Although linear regression models suggested a significant relationship between the behavioral and the Cochlea-Scan data in all the tested frequencies, the low R2 values in each model suggest an unexplained variability in the data. For the hearing-impaired group the mean difference in hearing levels at 1.5 and 2.0 kHz was below 5 dB, while at 4.0 kHz the behavioral threshold was overestimated by 7 dB. In this group the observed relationships were stronger (larger R[sup]2[/sup] values) than in the normal group, and the most significant relationships were observed at 2.0, 4.0, and 6.0 kHz (with descending R[sup]2[/sup] values).Conclusions:The data from this pilot study indicate that Cochlea-Scan assessment is possible and accurate in both normal-hearing and in hearing-impaired subjects. The margin of assessment in the hearing-impaired group was found to be closer to a PTA value of 40 dB and not 50 dB HL as previously suggested.
机译:背景:本研究的目的是评估使用新的声学协议(Cochlea-Scan)引起的听力阈值误差。材料/方法:本研究使用了35名正常听力和30名听力受损的成年人。听力阈值水平通过纯音测听(PTA)和Cochlea-Scan(Fischer-Zoth / Natus)DPOAE协议进行评估。结果:正常听力组在2.0、3.0和4.0 kHz时听力水平的平均差异低于8 dB HL。在1.5 kHz时,平均差约为9 dB HL,而在6.0 kHz时,行为阈值被高估了2 dB HL。尽管线性回归模型表明在所有测试频率下行为和Cochlea-Scan数据之间存在显着关系,但每个模型中的低R2值都表明数据存在无法解释的可变性。对于听力受损的人群,在1.5和2.0 kHz时听力水平的平均差异低于5 dB,而在4.0 kHz时,行为阈值被高估了7 dB。在该组中,观察到的关系比正常组中的关系更强(R [sup] 2 [/ sup]值更大),并且在2.0、4.0和6.0 kHz处观察到最显着的关系(R2下降) [/ sup]值。)结论:该初步研究的数据表明,对于正常听力和听力受损的受试者,Cochlea-Scan评估是可能且准确的。发现听力障碍组的评估余量更接近PTA值40 dB,而不是先前建议的HL 50 dB HL。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号