首页> 外文期刊>Medical science monitor : >Star GK Bileaflet Mechanical Valve Prosthesis-Patient Mismatch after Mitral Valve Replacement: A Chinese Multicenter Clinical Study
【24h】

Star GK Bileaflet Mechanical Valve Prosthesis-Patient Mismatch after Mitral Valve Replacement: A Chinese Multicenter Clinical Study

机译:Star GK Bileaflet二尖瓣置换术后机械瓣膜假体-患者不匹配:一项中国多中心临床研究

获取原文
           

摘要

BACKGROUND The aim of this study was to investigate the incidence and immediate and mid-term effects of heart valve prosthesis-patient mismatch (PPM) after mitral valve replacement using the GK bileaflet mechanical valve. MATERIAL AND METHODS A total of 493 cases of mechanical mitral valve replacement were performed in the departments of cardiac surgery in 7 hospitals from January 2000 to January 2008. The patients included 142 men and 351 women ages 21 to 67 (average age, 48.75). The patients were followed for 3 years after surgery. The effective orifice area index (EOAI), ≤1.2 cm2/m2, was detected during the follow-up period and was defined as PPM. The patients were assigned to either the PPM group or the non-PPM group. Finally, the preoperative, perioperative and postoperative indexes of the 2 groups of patients were compared. RESULTS A total of 157 patients had PPM 3 years after surgery. The incidence of PPM was 31.84%. Sixty-three patients in the PPM group received a 25-mm GK bileaflet valve (40.13%), 82 received a 27-mm valve (52.23%), and 12 (7.64%) received a 29-mm valve. There were significant differences in length of intensive care unit stay, duration of ventilator use, length of hospitalization, body surface area, EOAI, mean transmitral pressure gradient, and pulmonary artery pressure between the PPM and non-PPM group ([i]P[/i]0.05). CONCLUSIONS PPM after mitral valve replacement influences postoperative hemodynamics. Thus, larger-sized GK bileaflet mechanical valves are often used to reduce the risk of PPM.
机译:背景技术本研究的目的是研究使用GK双叶机械瓣膜置换二尖瓣后心脏瓣膜假体-患者不匹配(PPM)的发生率和近期和中期影响。材料与方法2000年1月至2008年1月,在7所医院的心脏外科部门共进行了493例机械性二尖瓣置换手术。患者包括142例男性和351例21至67岁的女性(平均年龄48.75)。术后随访3年。在随访期间检测到有效孔口面积指数(EOAI)≤1.2cm2 / m2,并定义为PPM。将患者分为PPM组或非PPM组。最后,比较了两组患者的术前,围术期和术后指标。结果共有157例患者术后3年发生了PPM。 PPM的发生率为31.84%。 PPM组中的63例患者接受了25 mm GK双叶瓣(40.13%),82例接受了27 mm瓣膜(52.23%),12例(7.64%)接受了29 mm瓣膜。在PPM组和非PPM组之间,重症监护病房住院时间,呼吸机使用时间,住院时间,体表面积,EOAI,平均透射压力梯度和肺动脉压力之间存在显着差异([i] P [ /i]0.05)。结论二尖瓣置换术后的PPM影响术后血液动力学。因此,通常使用较大尺寸的GK双叶机械瓣膜以降低PPM的风险。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号