Increased Sensitivity to Low Concentration Gadolinium Contrast by Optimized Heavily T2-weighted 3D-FLAIR to Visualize Endolymphatic Space

机译:通过优化T 2 加权的3D-FLAIR优化可视化的内淋巴空间,提高了对低浓度d对比度的敏感性

获取原文
       

摘要

Purpose: To increase the sensitivity of 3-dimensional fluid-attenuated inversion recovery (3D-FLAIR) to low concentration gadolinium (Gd)-based contrast medium, we optimized sequence parameters on a phantom and evaluated the optimized sequence in patients suspicious for endolymphatic hydrops. Materials and Methods: All scans were performed on a 3-tesla magnetic resonance (MR) unit using a 32-channel head coil. We optimized sequence parameters using a phantom filled with diluted Gd and compared the optimized protocol with 3D-FLAIR using conventional turbo spin echo sequence (3D-FLAIR-CONV). Nine patients underwent scanning using the newly optimized sequence and 3D-FLAIR-CONV 4 hours after double-dose administration of intravenous Gd. We subjectively scored separation of endo- and perilymph space and measured contrast-to-noise ratio (CNR) between endo- and perilymph. Results: The optimized sequence in the phantom study consisted of: repetition time, 9000 ms; echo time, 540 ms; inversion time, 2400 ms; low constant readout flip angle, 120° in the later part of the echo train. Image contrast became heavily T2-weighted (hT2W-3D-FLAIR). In patients, we recognized endolymphatic space for both the cochlea and vesti- bule significantly better by hT2W-3D-FLAIR than 3D-FLAIR-CONV ( P 2W-3D-FLAIR was more sensitive than the previous method to low concentration of Gd. Visualization of the endolymphatic space by double-dose administration of intravenous Gd would be more reliable using hT2W-3D-FLAIR.
机译:目的:为了提高3D液体衰减反转恢复(3D-FLAIR)对低浓度g(Gd)造影剂的敏感性,我们优化了人体模型上的序列参数,并评估了可疑内淋巴积液患者的最佳序列。材料和方法:所有扫描均使用32通道磁头线圈在3-tesla磁共振(MR)单元上进行。我们使用填充有稀释Gd的体模优化了序列参数,并使用常规涡轮自旋回波序列(3D-FLAIR-CONV)将优化后的协议与3D-FLAIR进行了比较。两次静脉注射Gd后4小时,使用新优化的序列和3D-FLAIR-CONV对9名患者进行了扫描。我们对内淋巴和外淋巴间隙进行主观评分,并测量内淋巴和外淋巴之间的对比噪声比(CNR)。结果:体模研究中优化的序列包括:重复时间9000 ms;回声时间540毫秒;反转时间2400毫秒;低恒定读数翻转角,在回波序列的后期,为120°。图像对比度变得很重T 2 加权(hT 2 W-3D-FLAIR)。在患者中,hT 2 W-3D-FLAIR比3D-FLAIR-CONV(P 2 W-3D-FLAIR)更好地识别了耳蜗和前庭的淋巴间隙比以前的方法对低浓度的Gd更为敏感,使用hT 2 W-3D-FLAIR双重剂量静脉注射Gd可以更直观地观察淋巴间隙。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号