A COMPARATIVE STUDY ON INSTRUMENTAL PRECISION OF REFRIGERATED AND NON-REFRIGERATED AUTO-ANALYZERS IN ORDER TO IMPROVE QUALITY ASSURANCE IN BIOCHEMISTRY LABORATORY

机译:冷藏和非冷藏自动分析仪的仪器精度按顺序提高生物化学实验室质量保证的比较研究

获取原文
获取外文期刊封面目录资料

摘要

Background and Objective: Quality control is one of the most important components in order to improve quality assurance in laboratories during analytical steps. For this purpose, coefficient of variation plays an important role. Due to the fast improvement in technology, application of inferential statistics for the comparisons of laboratory techniques, including instrument and chemicals and fast replacement of them would be technically essential. Materials and Methods: In this research study, we tried to determine the precision of one auto-analyzer with refrigerator and the other one without such facility for the analysis of glucose, triglyceride and cholesterol in 30 successive days and compared the coefficient of variations that are an important factor for analytical precision. Results: Comparing the means of aimed groups by paired t-test with P=0.05 in measurement of glucose, triglyceride and cholesterol showed that there was no significant difference between the two instruments. In determination of glucose, triglyceride and cholesterol by auto-analyzer with refrigerator, CV% of these analytes was 1.3%, 1.4%, and 0.9% respectively. By the auto-analyzer without refrigerator, the CV% of those analytes was 1.5%, 2.5% and 1.11% respectively. Conclusion: There is reasonably higher precision for auto-analyzer with refrigerator than the other one.
机译:背景与目的:质量控制是最重要的组成部分之一,以提高实验室在分析步骤中的质量保证。为此,变异系数起着重要的作用。由于技术的飞速发展,将推论统计学应用于实验室技术(包括仪器和化学药品)的比较中,并对其进行快速替换在技术上至关重要。材料和方法:在本研究中,我们试图确定一台带有冰箱的自动分析仪的精度,以及另一台没有这种设备的自动分析仪的精度,以连续30天分析葡萄糖,甘油三酸酯和胆固醇,并比较变异系数。分析精度的重要因素。结果:通过配对t检验和P = 0.05对目标人群进行血糖,甘油三酸酯和胆固醇测量的平均值进行比较,结果表明两种仪器之间没有显着差异。用冰箱自动分析仪测定葡萄糖,甘油三酸酯和胆固醇时,这些分析物的CV%分别为1.3%,1.4%和0.9%。通过不带冰箱的自动分析仪,这些分析物的CV%分别为1.5%,2.5%和1.11%。结论:带有冰箱的自动分析仪比其他自动分析仪具有更高的精度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号