Comparison of central corneal thickness estimated by an ultrasonic pachymeter and non-contact specular microscopy

机译:超声测厚仪和非接触式镜检显微镜估计的角膜中央厚度的比较

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Purpose: To compare central corneal thickness (CCT) measurements of healthy individuals obtained with ultrasonic pachymetry (UP) and non-contact specular microscopy (NCSM). Method: In total, 148 eyes of 74 subjects with no ocular or systemic diseases were included in the study. Central corneal thickness measurements of all patients performed with UP and NCCM were compared. Results: A total of 74 subjects (38 females) were included in this study. The mean age was 45.2 ?± 18.4 (range 12-85) years. The mean central corneal thickness of all 148 eyes was 546.9 ?± 40 ??m with UP and 510.8 ?± 42 ??m with NCSM. The mean central corneal thickness measured with NCSM was 35 ??m thinner than that measured with UP (p<0.001). A high degree of agreement was found between the two methods (r=0.942, p<0.001). Conclusions: Our results suggest that NCSM measures thinner corneas than UP and that the correction formula we identified should be applied when comparing between these two devices.
机译:目的:比较通过超声测厚法(UP)和非接触镜检显微镜(NCSM)获得的健康个体的中央角膜厚度(CCT)测量值。方法:本研究共纳入74名无眼或全身疾病的受试者的148只眼。比较了所有使用UP和NCCM进行的患者中央角膜厚度测量。结果:本研究共纳入74名受试者(38名女性)。平均年龄为45.2±18.4岁(范围12-85)。 UP的全部148只眼的中央角膜平均厚度为UP,NCSM的平均中央角膜厚度为546.9Ω±40Ωm,NCSM的平均角膜中央厚度为510.8Ω±42Ωm。用NCSM测量的平均中央角膜厚度比UP测量的厚度薄35 ?? m(p <0.001)。两种方法之间发现高度一致(r = 0.942,p <0.001)。结论:我们的结果表明,NCSM比UP可以测量更薄的角膜,并且在比较这两种设备时应采用我们确定的校正公式。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号