Purpose: of this paper was follow up of compatibility of femoral component taper from TiAl6V4 ELI acc.ISO5832-3 and ceramic head of the new type of total hip join replacement THR - ZRM.in vitro.Design/methodology/approach: from december 2006 to august 2007 we have tested THR taper component and ball ceramic head. On the base of our patents ( SK, CZ, EU, USA) from 1999 to 2006 we have designed and developed the new type of cementlesss THR - ZRM.. These total hip replacements were tested in our prospective study. We have evaluated compatibility and loading resistance of taper of THR ZRM.made from TiAl6V4alloy ELI and ceramic ball head Biolox.forte (28-12/14L) by tests of burst strength, pull out, rotational stability, fatigue and post-fatigue. Tests were monitored independent observer.Findings: in the burst test were tested 7 parts and average loading was 54 kN. The reached values comply with the requirements according to CeramTec procedure VA 02 04 4129 and the FDA Guidance document for the preparation of premarket notifications for ceramic ball hip systems. In the fatigue test all specimens test at a maximum load of 14 kN reached 10 million cycles without mechanical failure. The post fatigue test was performed with load of 46 kN and no failure of head and taper system was observed. The pull-off tests were made with pull out strength of 1439 N has been used on 5 parts and no pullout of ceramic ball head of taper femoral component was observed. Last was the test of rotational stability, where ball head was rotational stable under strength of 1254 Ncm.Research limitations/implications: this study was performed on the small group of spices but enough relevant according FDA and international ISO prescriptions.Originality/value: this study is original by publishing first results of the new cementless THR- ZRM.tests of compatibility and fatigue of metal taper and ceramic ball head. Value of this paper is in vitro testing and prepares ZRM.cementless THR for clinical use in human
机译:目的:本文针对TiAl6V4 ELI符合ISO5832-3的股骨组件锥形与新型全髋关节置换THR-ZRM陶瓷头的体外相容性设计/方法/方法:自2006年12月起到2007年8月,我们已经测试了THR锥度部件和球形陶瓷头。在1999年至2006年我们的专利(SK,CZ,欧盟,美国)的基础上,我们设计并开发了新型非骨水泥THR-ZRM。在我们的前瞻性研究中对这些全髋关节置换术进行了测试。我们通过测试爆破强度,拉出强度,旋转稳定性,疲劳性和疲劳后强度,评估了由TiAl6V4合金ELI和陶瓷球头Biolox.forte(28-12 / 14L)制成的THR ZRM锥度的相容性和耐载荷性。测试由独立的观察员进行监控。发现:在爆裂测试中测试了7个部件,平均负载为54 kN。达到的值符合根据CeramTec程序VA 02 04 4129和FDA指导文件为陶瓷球髋系统准备上市前通知的要求。在疲劳试验中,所有最大载荷为14 kN的试样在没有机械故障的情况下达到了1000万次循环。疲劳后测试的负荷为46 kN,未观察到头部和锥度系统的故障。在5个部件上使用1439N的拉拔强度进行拉拔试验,未观察到锥形股骨组件的陶瓷球头的拉拔。最后是旋转稳定性测试,其中球头在1254 Ncm的强度下是旋转稳定的。研究限制/意义:该研究是针对一小群香料进行的,但根据FDA和国际ISO处方具有足够的相关性。该研究是通过发表新型非水泥THR-ZRM的第一批结果而独创的。本文的价值是体外测试并制备ZRM.lessmental THR以供人类临床使用

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号