首页> 外文期刊>Analytical Sciences >Flow Injection Titration of Chloride in Electroplating Baths With a Silver Tubular Electrode as a Detector
【24h】

Flow Injection Titration of Chloride in Electroplating Baths With a Silver Tubular Electrode as a Detector

机译:镀银管电极检测电镀液中氯化物的流动注射滴定

获取原文
获取外文期刊封面目录资料

摘要

A flow injection system has been constructed for the "pseudo-titration" of chloride in electroplating baths with potentiometric detection. A silver ion-selective tubular electrode based on a homogeneous crystalline membrane was used. The FIA manifold has been developed and optimized in order to analyze samples with a chloride concentration range of 6.O×10-2M to 2.5M. The proposed method is precise (relative standard deviation about 5%) and the results obtained by a simultaneous application of the FIA method and an automatic precipitation titration method, adopted as a reference technique, are in good agreement (r=0.996). The sampling rate for the FIA manifold is 78 samples per hour.
机译:已经构建了一种流动注射系统,用于电位滴定法对电镀液中的氯化物进行“伪滴定”。使用基于均质结晶膜的银离子选择性管状电极。已开发并优化了FIA歧管,以分析氯化物浓度范围为6.O×10-2M至2.5M的样品。所提出的方法是精确的(相对标准偏差约为5%),同时采用FIA方法和自动沉淀滴定法(作为参考技术)获得的结果吻合良好(r = 0.996)。 FIA歧管的采样率为每小时78个样本。