首页> 外文期刊>Analytical Sciences >Accurate Determination of Trace Molybdenum in Drinking Water by Isotope Dilution Inductively Coupled Plasma Mass Spectrometry
【24h】

Accurate Determination of Trace Molybdenum in Drinking Water by Isotope Dilution Inductively Coupled Plasma Mass Spectrometry

机译:同位素稀释电感耦合等离子体质谱法准确测定饮用水中痕量钼

获取原文
           

摘要

A method for accurate and precise determination of trace molybdenum in drinking water by isotope dilution inductively coupled plasma mass spectrometry was developed, given the concentrations of Mo in drinking-water samples from Chaoyang and Changping districts of Beijing (China) as 1.017 ± 0.008 and 1.033 ± 0.007 μg kg~(?1) (k = 2), respectively. Special care was taken for the validation of the proposed ID-ICPMS method using CRM 7203-a, a certified reference material for elemental analysis of tap water.
机译:提出了一种以同位素稀释电感耦合等离子体质谱法准确,准确地测定饮用水中痕量钼的方法,假定北京朝阳区和昌平区饮用水样品中的钼浓度为1.017±0.008和1.033分别为±0.007μgkg·(-1)(k = 2)。使用CRM 7203-a(一种用于自来水元素分析的认证参考材料)对提出的ID-ICPMS方法进行验证时要格外小心。