Fast Simultaneous Determination of Niobium and Tantalum by Kalman Filter Analysis with Flow Injection Chemiluminescence Method

机译:流动注射化学发光法卡尔曼滤波快速快速测定铌和钽。

获取原文
           

摘要

A fast and highly efficient Kalman Filter analysis-flow injection chemiluminescence (FI-CL) method was developed to simultaneously determine trace amounts of niobium and tantalum in geological samples. The method, without the boring process of separation and dear instruments, is suitable for field scene analysis. The mixed chemiluminescence kinetic curve was analyzed by a Kalman Filter (KF) in this method to realize the simultaneous determination of niobium and tantalum. Possible interference elements in the determination were investigated. Under the selected conditions, the detection limits (3 σ, n = 11) of niobium(V) and tantalum(V) were 2.1 × 10-3 µg g-1 and 4.0 × 10-3 µg g-1, respectively, and the relative standard deviations were 4.9% and 3.3% (n = 9). The method was applied to the determination of niobium and tantalum in geological samples with satisfactory results.
机译:建立了一种快速高效的卡尔曼滤波器分析-流动注射化学发光(FI-CL)方法,可同时测定地质样品中的痕量铌和钽。该方法无需费时费力的分离和昂贵的仪器操作,适用于现场场景分析。通过卡尔曼滤波器(KF)分析混合化学发光动力学曲线,可同时测定铌和钽。研究了测定中可能的干扰元素。在选定的条件下,铌(V)和钽(V)的检出限(3σ,n = 11)分别为2.1×10-3 µg g-1和4.0×10-3 µg g-1,并且相对标准偏差为4.9%和3.3%(n = 9)。该方法用于地质样品中铌和钽的测定,结果令人满意。