The limitations of ion-selective electrodes (ISEs) presently in use mainly concern the stability of the calibration values. By applying optimal membrane technologies and optimized reference electrodes a perfectly symmetric measuring system including a symmetric ISE membrane has been developed. Thus, stable standard potentials of the cell assembly, both before and after serum contact, as well as theoretical slopes of the electrode response have been obtained. Results of calibration-free assays of K+ and Na+ concentrations in aqueous calibration solutions and in undiluted human serum are presented. The precision required in fully met for K+ measurements in aqueous solutions and in sera of the normal as well as pathological concentration range. In addition, the precision of Na+ assays in aqueous calibration solutions and in serum samples of the normal concentration range is close to the required values.
机译:当前使用的离子选择电极(ISE)的局限性主要涉及校准值的稳定性。通过应用最佳膜技术和优化的参比电极,已开发出包括对称ISE膜的完美对称测量系统。因此,已经获得了在血清接触之前和之后细胞组件的稳定标准电位,以及电极响应的理论斜率。给出了水校准溶液和未稀释人血清中K +和Na +浓度的无校准测定结果。完全满足水溶液和血清中K +的正常测量以及病理浓度范围所需的精度。另外,在水溶液校准溶液和正常浓度范围的血清样品中,Na +分析的精度接近要求的值。