Evaluation of the Effectiveness of Dental Curing Light in Dental Offices Located in Karaj in 2016

机译:2016年位于Karaj的牙科诊所的牙科光固化机效果评估

获取原文
           

摘要

Objective : Light-cured composites and restorative materials are widely used in dentistry. The success of restoration largely depends on the effectiveness of the curing light, including the duration of irradiation and wavelength. This study was conducted to evaluate the effectiveness of curing light devices routinely used in dental offices of Karaj in 2016. Methods: In this descriptive-analytical study, 120 QTH and LED curing light devices used in Karaj dental offices were randomly evaluated. A standard questionnaire was used to collect the variables of the device type, age, and brand, number of repairs, reasons for repair, time of last repair, number of bulb replacements, time of last bulb replacement, device being fitted with a measure severity, duration of irradiation for each composite layer, frequency of using the equipment in a day, sterilization method, light intensity, and number of office days in a week. Data were analyzed with SPSS using Spearman correlation coefficient, independent t test, and chi-square at a level of 0.05%. Results: The results showed that the light intensity of all devices was above 200 mw/cmsup2/sup. The light intensity of 11.1% of QTH devices was unfavorable while the light intensity of all LED devices was favorable. The results showed a significant correlation between brand and light intensity, between the number of bulb replacement and device type, between the number of bulb replacements and light intensity, between the number of bulb replacements and clinical age of the device, and between the number of repairs and the device type (P≤0.05). There was no significant relationship between the device type and repair reason. The results also showed an indirect relationship between the clinical age of the device and light intensity. A significant difference was found in light intensity between QTH and LED devices. Conclusion: The light intensity of 11.1% of QTH devices was unfavorable, and the main reason for repair was tip breakage. Light intensity decreased with an increase in the clinical age of the device; therefore, regular quality control and timely bulb replacement are of great importance.
机译:目的:光固化复合材料和修复材料被广泛用于牙科。恢复的成功很大程度上取决于固化光的有效性,包括照射的持续时间和波长。这项研究旨在评估Karaj牙科诊所在2016年常规使用的固化灯设备的有效性。方法:在描述性分析研究中,随机评估了Karaj牙科诊所使用的120 QTH和LED固化灯设备。使用标准调查表收集设备类型,年龄和品牌,维修次数,维修原因,最后维修时间,灯泡更换次数,灯泡最后更换时间,设备是否符合测量严重性的变量,每个复合层的照射时间,一天使用设备的频率,灭菌方法,光强度以及一周的办公天数。使用Spearman相关系数,独立t检验和卡方检验,采用SPSS分析数据,水平为0.05%。结果:结果表明,所有器件的光强度均高于200 mw / cm 2 。 QTH器件的11.1%的光强度是不利的,而所有LED器件的光强度都是良好的。结果显示品牌与光强度之间,灯泡更换数量与设备类型之间,灯泡更换数量与光源强度之间,灯泡更换数量与设备临床年龄之间以及灯泡数量之间存在显着相关性。维修和设备类型(P≤0.05)。设备类型和维修原因之间没有显着关系。结果还显示了该装置的临床年龄和光强度之间的间接关系。 QTH和LED器件之间的光强度存在显着差异。结论:QTH装置的11.1%的光强度是不利的,并且维修的主要原因是尖端断裂。随着设备临床年龄的增加,光强度降低。因此,定期的质量控制和及时更换灯泡至关重要。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号