首页> 外文期刊>Advances in Physics Theories and Applications >Quality Estimation of Filtration of Diagnostic X-Ray at Federal Medical Centre and Bishop Murray Hospital Makurdi using Half-Value Layer (HVL)
【24h】

Quality Estimation of Filtration of Diagnostic X-Ray at Federal Medical Centre and Bishop Murray Hospital Makurdi using Half-Value Layer (HVL)

机译:使用半值层(HVL)在联邦医疗中心和Bishop Murray医院Makurdi进行诊断X射线过滤的质量评估

获取原文
           

摘要

Quality filtration of diagnostic X-rays in use at Federal Medical Centre and Bishop Murray Hospital Makurdi were checked using half-value layer (HVL) at 80kVp, 70kVp and 60kVp. A simple technique of determining HVL which is based on reducing the intensity of X-ray beam to half its original value using aluminium filters added at 2cm from the table and dose detected using DIAVOLT placed at 98cm from the centre of the X-ray tubes. Results were obtained from the plot of ln Dose (mGy) Versus Aluminium thickness (mm) where the slope gives the attenuation coefficient (m). The attenuation coefficient was then used in calculating HVL using the relationship HVL=In2/m. HVL values calculated were then compared with the minimum acceptable HVL values at the kVp setting as recommended by the international Commission on Radiological units and measurement.
机译:使用80kVp,70kVp和60kVp的半值层(HVL)检查了联邦医疗中心和Bishop Murray医院Makurdi使用的诊断X射线的质量过滤。一种简单的确定HVL的技术,其基础是,使用距桌子2cm处添加的铝滤光片,将X射线束的强度降低至其原始值的一半,并使用距X射线管中心98cm处的DIAVOLT检测剂量。从ln剂量(mGy)对铝厚度(mm)的曲线图获得了结果,其中斜率给出了衰减系数(m)。然后使用关系式HVL = In2 / m将衰减系数用于计算HVL。然后将计算的HVL值与国际放射单位和测量委员会建议的kVp设置下的最小可接受HVL值进行比较。