Measurement of migration of a humeral head resurfacing prosthesis using radiostereometry without implant marking

机译:不使用植入物标记的放射立体法测量肱骨头表面修复假体的迁移

获取原文
           

摘要

Introduction Standard radiostereometric analysis of prosthetic migration requires that tantalum beads are inserted into the implant. For manufacturing reasons, this is not possible for humeral head resurfacing implants. We therefore used marker-free radiostereometry, developed for metal-backed acetabular cups, on a dummy model to validate the method for a humeral head resurfacing prosthesis.Material and methods 3 hemispherical resurfacing prostheses of different sizes were marked with tantalum beads and mounted in a sawbone. Standard and marker-free radiostereometry was then done repeatedly with gradual shifts of position of the prosthesis between each analysis. The marker-free algorithm was then compared to the standard to determine the accuracy.Results The accuracy for marker-free radiostereometry was 0.22–0.47 mm for translations and 0.92–1.56 degrees for rotations.Interpretation Based on our results, marker-free radiostereometry can be used to measure migration of humeral head resurfacing prostheses. ...
机译:简介义肢迁移的标准放射线立体分析需要将钽珠插入植入物中。出于制造原因,这对于肱骨头表面置换植入物是不可能的。因此,我们在假人模型上使用了针对金属支撑髋臼杯开发的无标记放射线立体测量法,以验证肱骨头表面置换假体的方法。材料和方法3种不同尺寸的半球形表面置换假体用钽珠标记并安装在锯骨。然后,在每次分析之间假体位置逐渐移动的情况下,重复进行标准的无标记放射线立体测量。然后将无标记的算法与标准算法进行比较以确定结果的准确性。结果无标记的放射线立体测量的精度为0.22-0.47 mm,旋转的精度为0.92-1.56度。用于测量肱骨头表面修复假体的迁移情况。 ...

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号