首页> 外文期刊>Acta orthopaedica. >Measurement of migration of a humeral head resurfacing prosthesis using radiostereometry without implant marking: an experimental study.
【24h】

Measurement of migration of a humeral head resurfacing prosthesis using radiostereometry without implant marking: an experimental study.

机译:使用不带植入物标记的放射立体测量法测量肱骨头表面修复假体的迁移:一项实验研究。

获取原文
获取原文并翻译 | 示例
           

摘要

INTRODUCTION: Standard radiostereometric analysis of prosthetic migration requires that tantalum beads are inserted into the implant. For manufacturing reasons, this is not possible for humeral head resurfacing implants. We therefore used marker-free radiostereometry, developed for metal-backed acetabular cups, on a dummy model to validate the method for a humeral head resurfacing prosthesis. MATERIAL AND METHODS: 3 hemispherical resurfacing prostheses of different sizes were marked with tantalum beads and mounted in a sawbone. Standard and marker-free radiostereometry was then done repeatedly with gradual shifts of position of the prosthesis between each analysis. The marker-free algorithm was then compared to the standard to determine the accuracy. RESULTS: The accuracy for marker-free radiostereometry was 0.22-0.47 mm for translations and 0.92-1.56 degrees for rotations. INTERPRETATION: Based on our results, marker-free radiostereometry can be used to measure migration of humeral head resurfacing prostheses. This indicates that implant marking is not required when doing radiostereometry on humeral head resurfacing in clinical trials.
机译:引言:对假体迁移进行标准放射线立体分析需要将钽珠插入植入物中。出于制造原因,这对于肱骨头表面置换植入物是不可能的。因此,我们在虚拟模型上使用了针对金属支撑髋臼杯开发的无标记放射线立体测量法,以验证肱骨头表面修复假体的方法。材料与方法:3个不同大小的半球形修复体用钽珠标记并安装在锯骨中。然后,在每次分析之间假体位置逐渐变化的情况下,重复进行标准和无标记的放射立体测量。然后将无标记算法与标准算法进行比较以确定准确性。结果:无标记放射性立体测量的平移精度为0.22-0.47 mm,旋转精度为0.92-1.56度。解释:根据我们的结果,无标记放射线立体照相术可用于测量肱骨头表面修复假体的迁移。这表明在临床试验中对肱骨头置换进行放射立体测量时不需要植入物标记。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号