首页> 外文期刊>Acta orthopaedica. >Migration analysis of a metaphyseal anchored short-stem hip prosthesis
【24h】

Migration analysis of a metaphyseal anchored short-stem hip prosthesis

机译:干phy端短柄髋关节假体的迁移分析

获取原文
获取外文期刊封面目录资料

摘要

Background and purpose Metaphyseal anchored short-stem hip implants were designed to improve load transmission and preserve femoral bone stock. Until now, only few outcome data have been available and migration studies are one of the few ways of obtaining data that are predictive of implant survival. We therefore evaluated a metaphyseal anchored short-stem hip implant by Ein Bild Roentgen Analyse femoral component analysis (EBRA-FCA). Patients and methods First, the EBRA-FCA method was validated for the short-stem hip implant. Then 80 of the first 100 consecutive implants were evaluated after at least 2 years. Clinical assessment was performed using the WOMAC and the UCLA score. Results After 2.7 (2.0–4.2), years none of the implants had been revised and by that time the stems had subsided by a mean of 0.7 mm (SD 1.8) (95% CI: 0.3–1.1). Of the 80 implants, 78 were stable after 2 years, with 74 being primary stable and 4 showing secondary stabilization after initial subsidence. Continuous migration was see...
机译:背景和目的干phy端锚固的短柄髋关节植入物旨在改善负荷传递并保留股骨骨量。到目前为止,只有很少的结果数据可用,并且迁移研究是获得可预测植入物存活率的数据的几种方法之一。因此,我们通过Ein Bild Roentgen分析股骨成分分析(EBRA-FCA)评估了干phy端短柄髋关节植入物。患者和方法首先,EBRA-FCA方法已被验证用于短柄髋关节植入物。然后至少两年后评估前100个连续植入物中的80个。使用WOMAC和UCLA评分进行临床评估。结果2.7(2.0–4.2)岁后,多年没有任何植入物被修整,到那时茎平均平移了0.7毫米(SD 1.8)(95%CI:0.3–1.1)。在80种植入物中,有2种在2年后保持稳定,其中74种保持了最初的稳定性,有4种表现出了最初的沉降后的次要稳定性。看到持续迁移...

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号