In vivo and ex vivo measurement of polyethylene wear in total hip arthroplasty

机译:全髋关节置换术中聚乙烯磨损的体内和离体测量

获取原文

摘要

Background — Determination of the amount of wear in a polyethylene liner following total hip arthroplasty (THA) is important for both the clinical care of individual patients and the development of new types of liners. Patients and methods — We measured in vivo wear of the polyethylene liner using computed tomography (CT) (obtained in the course of regular clinical care) and compared it to coordinate-measuring machine (CMM) readings. Also, changes in liner thickness of the same retrieved polyethylene liner were measured using a micrometer, and were compared to CT and CMM measurements. The distance between the centers of the acetabular cup and femoral head component was measured in 3D CT, using a semi-automatic analysis method. CMM readings were performed on each acetabular liner and data were analyzed using 3D computer-aided design software. Micrometer readings compared the thickest and thinnest regions of the liner. We analyzed 10 THA CTs and retrievals that met minimal requirements for CT slice thicknes...
机译:背景技术–确定全髋关节置换术(THA)后聚乙烯衬里的磨损量对于单个患者的临床护理以及新型衬里的开发都很重要。患者和方法-我们使用计算机断层扫描(CT)(在常规临床护理过程中获得)测量了聚乙烯衬里的体内磨损,并将其与坐标测量机(CMM)读数进行了比较。同样,使用千分尺测量相同回收的聚乙烯衬里的衬里厚度的变化,并将其与CT和CMM测量值进行比较。使用半自动分析方法在3D CT中测量髋臼杯中心与股骨头组件之间的距离。在每个髋臼内衬上进行CMM读数,并使用3D计算机辅助设计软件分析数据。千分尺读数比较了衬管的最厚和最薄区域。我们分析了10个THA CT和满足CT切片厚度最低要求的取回物...